29
Using the iSleep 20
iSleep 20 clinician’s manual
Doc. 003923 En-Us V-2
5.6 Using an External Battery
The iSleep 20 can be powered from:
• The Breas external EB 2 Battery Pack.
• A 12 V external power source. The iSleep DC/DC converter 12-24 V needs
to be used.
• A 24 V external power source. The iSleep battery cable needs to be used.
Battery Operation Time
The operation time is dependent on the battery condition, its capacity, the ambi-
ent air temperature and the iSleep 20 pressure setting.
The operation time is based on the following settings:
1
Connect the external DC cable to
the iSleep 20. Make sure it is fitted
correctly.
2
Connect the other end of the cable to the battery source.
• Always make a function check to test the battery condition before operat-
ing the iSleep 20 from an external battery source.
• The battery must be disconnected when the iSleep 20 is switched off, oth-
erwise the battery will be discharged.
S
ETTING
V
ALUE
Humidifier setting
5 or Off
CPAP pressure
10 cmH
2
O
E
NVIRONMENTAL
C
ONDITIONS
Ambient temperature
68°F (20°C)
Altitude
Sea level
Battery status
New battery fully charged