11
Safety Information
iSleep 20 clinician’s manual
Doc. 003923 En-Us V-2
2.5 Usage of Filters
2.6 Cleaning and Maintenance
• Always use the iSleep 20 with a patient air inlet filter installed. Only use fil-
ters that are specified in this manual.
• Replace or clean the filters regularly to ensure correct function of the
iSleep 20, especially when changing patient. Failure to replace or clean a
dirty filter may cause the iSleep 20 to operate at higher temperatures than
intended.
• When operating the iSleep 20, make sure that the air inlet and the filter is
not obstructed or occluded.
• If the iSleep 20 is used in a clinic by several patients, a low resistance bacte-
ria filter is recommended between the air outlet and the patient circuit to
prevent patient cross-contamination. Reuse of mask or bacteria filter may
expose patients to contagious agents.
• Do not connect any filter to the HA 01 humidifier.
• The iSleep 20 shall be cleaned and maintained in accordance with this
clini-
cian’s
manual.
• Do not attempt to autoclave or sterilise the iSleep 20.
• The iSleep 20 shall be subjected to maintenance, service and control and
any applicable upgrades, in accordance with Breas service instructions.
• The iSleep 20 should only be repaired or modified in accordance with
Breas service manuals, technical bulletins, and any special service instruc-
tions, by service technicians authorized by Breas Medical AB.
• Do not under any circumstances attempt to service or repair the iSleep 20
yourself. If you do so, the manufacturer will no longer be responsible for
the performance and safety of the iSleep 20.