15 Maintenance Instructions
NIPPY Clearway 2 Service Manual
2210.2657 Rev 3.0
Internal inspection
WARNING!
Disconnect the AC mains supply before opening the Clearway 2
case
Observe ESD precautions when handling Printed Circuit Boards
Every 5 years or following a repair, perform the 12 month checks plus an
internal inspection detailed in this chapter.
Open the case as described in this manual.
Internal clean
Clean out any dust which may have accumulated inside the Clearway 2.
Inspect electrical connections
Visually inspect all internal electrical cables. Cable sheathing must be intact and
in good condition and ensure cables have not been damaged. Verify that all
connections are securely made.
Inspect mechanical fastenings
Ensure 5 x command/display PCB fixings are secure (M3 x 8)
Ensure 2 x front switch fixings are secure (M3 x 8)
Ensure 4 x power supply fixings are secure (M3 x 8)
Ensure 4 x valve assembly fixings are secure (M4 x 10)
Ensure 2 x fixings for rigid pipe are secure (M3 x 8)
Check cooling fan is fitted securely.
Check blower jacket is securely closed.
Check silicone pipes are all securely fitted to blower and valve.
Check pressure measurement pipes are securely connected.
Содержание Clearway 2
Страница 1: ...Service Manual Technical Instructions for Use 2210 2657 Revision 3 0...
Страница 25: ...25 Part identification NIPPY Clearway 2 Service Manual 2210 2657 Rev 3 0 Display PCB...
Страница 26: ...Part identification 26 2210 2657 Rev 3 0 NIPPY Clearway 2 Service Manual Inside Rear Case...
Страница 27: ...27 Part identification NIPPY Clearway 2 Service Manual 2210 2657 Rev 3 0 Power PCB...
Страница 30: ...Functional Diagrams 30 2210 2657 Rev 3 0 NIPPY Clearway 2 Service Manual Control PCB block diagram...
Страница 31: ...31 Functional Diagrams NIPPY Clearway 2 Service Manual 2210 2657 Rev 3 0 Display PCB block diagram...
Страница 32: ...Functional Diagrams 32 2210 2657 Rev 3 0 NIPPY Clearway 2 Service Manual Pneumatic diagrams...