84 How to Use the Vivo 45
Vivo 45 User Manual
Doc.006149 En Y-2
Storing the Internal Battery and the Click-in Battery
Storage longer than 1 month should be initiated with half-charged batter-
ies in order to maintain maximum capacity. Optimal storage temperature
is 5 to 30°C (41 to 86°F).
External DC
The Vivo 45 can be operated from:
• A 12 V external DC source using the 12/24 V converter.
• A 24 V external DC source using the external battery cable.
• Both AC power supply and external DC using the Y-cable.
With both AC power supply and external DC source connected, the
Vivo 45 will automatically switch over to the external DC source if the
AC power supply fails. The external DC voltage level is shown under
“Others”, “Device Information” in the menu.
BATTERY
O
PERATING
T
IME
Internal Battery
2.5 h
Click-in Battery
6.5 h
Do not connect the Vivo 45 to a wheelchair unless the operating manual
for the wheelchair permits this as this can affect the Vivo 45 performance
and consequently result in patient death.
Only use a battery charger compliant to IEC 60601-1 if you are charging a
battery that at the same time is connected to the Vivo 45.
1
Connect the external DC cable
to the Vivo 45. Make sure that it is
fitted correctly.
2
Connect the other end of the cable to the DC source.
Содержание Vivo 45
Страница 10: ...10 Introduction Vivo 45 User Manual Doc 006149 En Y 2...
Страница 131: ...131 Alarms Vivo 45 User Manual Doc 006149 En Y 2...
Страница 176: ...176 Cleaning and Maintenance Vivo 45 User Manual Doc 006149 En Y 2...
Страница 196: ...196 Technical Specifications Vivo 45 User Manual Doc 006149 En Y 2...
Страница 206: ...206 Accessories and Parts Vivo 45 User Manual Doc 006149 En Y 2...
Страница 208: ...208 Patient Settings Vivo 45 User Manual Doc 006149 En Y 2...
Страница 210: ...210 FAA Compliance Vivo 45 User Manual Doc 006149 En Y 2...
Страница 226: ...226 Index Vivo 45 User Manual Doc 006149 En Y 2...