![Breas Medical Vivo 45 Скачать руководство пользователя страница 83](http://html1.mh-extra.com/html/breas-medical/vivo-45/vivo-45_user-manual_2806367083.webp)
83
How to Use the Vivo 45
Vivo 45 User Manual
Doc.006149 En Y-2
Battery Operating Time (Internal and Click-in)
The operation time is dependent on the battery condition, its capacity,
the ambient air temperature and the Vivo 45 pressure setting. These data
are based on new and fully charged batteries.
*: These ventilator settings affect the operating time significantly.
P
ARAMETER
E
XAMPLE
E
NVIRONMENTAL
C
ONDITIONS
Ambient temperature
20°C (68°F)
V
ENTILATOR
SETTINGS
Mode
PCV
Pressure*
20 cmH
2
O
PEEP
4 cmH
2
O
Breath Rate*
20 bpm
Insp. Time*
1.0 s
I:E
1:2
Insp. Trigger
Off
Rise Time
1
Target Volume
Off
Display Light*
Off
Light Intensity*
-
M
ONITORED
V
ALUE
Tidal Volume
800 ml
Resistance
5 hPa (l/s)
-1
Compliance
50 ml (hPa)
-1
Содержание Vivo 45
Страница 10: ...10 Introduction Vivo 45 User Manual Doc 006149 En Y 2...
Страница 131: ...131 Alarms Vivo 45 User Manual Doc 006149 En Y 2...
Страница 176: ...176 Cleaning and Maintenance Vivo 45 User Manual Doc 006149 En Y 2...
Страница 196: ...196 Technical Specifications Vivo 45 User Manual Doc 006149 En Y 2...
Страница 206: ...206 Accessories and Parts Vivo 45 User Manual Doc 006149 En Y 2...
Страница 208: ...208 Patient Settings Vivo 45 User Manual Doc 006149 En Y 2...
Страница 210: ...210 FAA Compliance Vivo 45 User Manual Doc 006149 En Y 2...
Страница 226: ...226 Index Vivo 45 User Manual Doc 006149 En Y 2...