How to Use the Vivo 45
Vivo 45 User Manual
Doc.006149 En Y-2
delivered volume and the set target volume. Automatic pressure adjust-
ments will be made in between two adjustable limits (min pressure and
max pressure) in order to deliver the optimal support to the patient.
PCV(A+TgV) – Assisted Pressure Controlled Ventilation with Tar-
get Volume
The PCV(A+TgV) mode acts like the PCV(A) mode, but with an added
regulation of the pressure. Target volume is a feature that automatically
adapts the pressure to make sure that the Vivo 45 delivers the desired set
target volume to the patient. The delivered volume is compared to the set
target volume on a breath by breath basis. The delivered pressure for the
next breath will be increased or decreased depending on the difference
between the delivered volume and the set target volume. Automatic pres-
sure adjustments will be made in between two adjustable limits (min pres-
sure and max pressure) in order to deliver the optimal support to the
patient.
CPAP – Continuous Positive Airway Pressure
In CPAP mode the Vivo 45 is applying a continuous positive pressure to
the airways. The flow will automatically be adjusted to maintain the set
CPAP level.
See “Target Volume” on page 68 for more information about Target Volume.
See “Target Volume” on page 68 for more information about Target Volume.
Содержание Vivo 45
Страница 10: ...10 Introduction Vivo 45 User Manual Doc 006149 En Y 2...
Страница 131: ...131 Alarms Vivo 45 User Manual Doc 006149 En Y 2...
Страница 176: ...176 Cleaning and Maintenance Vivo 45 User Manual Doc 006149 En Y 2...
Страница 196: ...196 Technical Specifications Vivo 45 User Manual Doc 006149 En Y 2...
Страница 206: ...206 Accessories and Parts Vivo 45 User Manual Doc 006149 En Y 2...
Страница 208: ...208 Patient Settings Vivo 45 User Manual Doc 006149 En Y 2...
Страница 210: ...210 FAA Compliance Vivo 45 User Manual Doc 006149 En Y 2...
Страница 226: ...226 Index Vivo 45 User Manual Doc 006149 En Y 2...