Cleaning and Maintenance
Vivo 45 User Manual
Doc.006149 En Y-2
Air Pathway Disinfection
The table below lists the parts that might get contaminated by exhaled
gases or bodily fluids during normal use or single fault condition.
In case of contamination, the internal air pathways of the Vivo 45 may be
disinfected up to 10 times by a maximum 60 minute long validated ozone
gas process.
C
ONDITION
P
ARTS
With bacteria filter
Patient circuit
Bacteria filter
Without bacteria filter
Patient circuit
FiO
2
sensor (if used)
Patient air outlet
Pneumatic unit
Air bypass unit/water chamber
Blower
Inlet silencer
Air inlet with filters