170 Cleaning and Maintenance
Vivo 45 User Manual
Doc.006149 En Y-2
7.1 Cleaning the Vivo 45
Main Unit
To avoid electrical shock, disconnect the power supply to the Vivo 45
before cleaning. Do not immerse the Vivo 45 into any fluids.
• Always be careful when cleaning to ensure that you do not damage
any equipment.
• Fluid must not be allowed to enter the Vivo 45.
• Never apply any liquids directly on the Vivo 45 by spraying, splashing
or pouring. Use a moistened lint-free cloth when cleaning.
• Do not use an excessive amount of liquid when cleaning the Vivo 45.
• Do not autoclave the Vivo 45.
1
Switch off the Vivo 45 and disconnect the power supply.
2
Remove the patient circuit.
3
Disconnect all electric cables.
4
Clean the outside of the Vivo 45 using a lint-free cloth with a mild
soap solution, and/or ethanol 70% for surface disinfection.
5
If the click-in humidifier is used, clean it as described in “Cleaning
the Water Chamber” on page 103.
6
Reconnect the patient circuit. Make sure all parts are dry before the
Vivo 45 is put into operation.
Содержание Vivo 45
Страница 10: ...10 Introduction Vivo 45 User Manual Doc 006149 En Y 2...
Страница 131: ...131 Alarms Vivo 45 User Manual Doc 006149 En Y 2...
Страница 176: ...176 Cleaning and Maintenance Vivo 45 User Manual Doc 006149 En Y 2...
Страница 196: ...196 Technical Specifications Vivo 45 User Manual Doc 006149 En Y 2...
Страница 206: ...206 Accessories and Parts Vivo 45 User Manual Doc 006149 En Y 2...
Страница 208: ...208 Patient Settings Vivo 45 User Manual Doc 006149 En Y 2...
Страница 210: ...210 FAA Compliance Vivo 45 User Manual Doc 006149 En Y 2...
Страница 226: ...226 Index Vivo 45 User Manual Doc 006149 En Y 2...