Alarms
Vivo 45 User Manual
Doc.006149 En Y-2
CO
2
Sensor Failure/Disconnection Alarm (CO
2
Discon-
nected)
I
TEM
D
ESCRIPTION
Definition
A CO
2
Sensor Failure/Disconnection alarm
will be given when communication between
the Vivo 45 and the CO
2
sensor has been lost.
Check the CO
2
sensor.
Priority
High
Possible cause
• CO
2
Sensor disconnected.
• Failure in the CO
2
sensor.
Ventilator action
The Vivo 45 will continue treatment with the
same settings.
Indication