High Pulse Rate Alarm
I
TEM
D
ESCRIPTION
Definition
A High Pulse Rate alarm will be given when the
measured pulse rate exceeds the alarm limit for
15 seconds.
Priority
Medium
Possible cause
• Insufficient ventilatory support.
• FiO
2
too low.
• EPAP too high.
Setting min
20 bpm
Setting max
250 bpm, Off
Setting resolution
5 bpm
Ventilator action
The Vivo 45 will continue treatment with the
same settings.
Indication