Alarms
Vivo 45 User Manual
Doc.006149 En Y-2
Low SpO
2
Alarm
I
TEM
D
ESCRIPTION
Definition
A Low SpO
2
alarm will be given when the
measured SpO
2
is below the alarm limit for
30 seconds.
Priority
High
Possible cause
• Added O2 flow is too low.
• Oxygen inlet is disconnected.
• Delivered tidal volumes are too small.
Setting min
Off, 85%
Setting max
100%
Setting resolution
1%
Ventilator action
The Vivo 45 will continue treatment with
the same settings.
Indication