![Breas Medical Vivo 45 Скачать руководство пользователя страница 120](http://html1.mh-extra.com/html/breas-medical/vivo-45/vivo-45_user-manual_2806367120.webp)
120 Alarms
Vivo 45 User Manual
Doc.006149 En Y-2
Rebreathing
I
TEM
D
ESCRIPTION
Definition
• A Rebreathing alarm will be given when the
measured leakage is lower than the expected
leakage flow at the set pressure for more than
15 ±0.5 seconds.
Priority
High
Possible cause
Obstructed or occluded leakage port.
Setting range
On, Off
Ventilator action
The Vivo 45 will continue treatment with the
same settings.
Indication