112 Alarms
Vivo 45 User Manual
Doc.006149 En Y-2
High Expired Tidal Volume Alarm (High Vte)
I
TEM
D
ESCRIPTION
Definition
A High Expired Tidal Volume alarm will be
given when the monitored Expired Tidal Vol-
ume exceeds the set limit for the High Expired
Tidal Volume alarm for 15 ±0.5 seconds.
Priority
Medium
Possible cause
• Mismatch between Expired Tidal Volume and
alarm setting.
• Pressure settings causing the Expired Tidal
Volume to exceed the set alarm level.
Setting min
150 ml (Adult), 20 ml (Paediatric)
Setting max
2500 ml, Off (Adult),
600 ml, Off (Paediatric)
Setting resolution
10 below 600 ml, 100 above 600 ml
Ventilator action
The Vivo 45 will continue treatment with the
same settings.
Indication