Perfusor® compact, 2.1 gb   

8 -   1

8

8 -

Procedural Instructions on the TSC

Visual Inspection 

Unit, in General

Completeness, external damage, safe fit of the battery
compartment cover and syringe table.
Check cleanliness of device. Check labels and readability.

Syringe fastening

Check function with OPS 50 ml syringe.
(Syringe holder, axial positioner, drive head, clamp, and push-
button sensor)

Membrane keyboard

Check adhesion, cleanliness and fit.

Battery compartment cover and battery contacts

Check state of contacts (tight fit, not bent).

Unit feet

Check unit feet for completeness and proper fit.

Mains Power Lead and Connector

Completeness, damage.

MFC Lead and Connector

Completeness, damage.

Holder for pole fixation, side snap-in mechanism

Check function.

8 - 6

2.1

Содержание perfusor compact

Страница 1: ...Perfusor compact Service Manual 0 10 3 1 0 Version 3 1 English...

Страница 2: ...pact English 8713 9112 Languages of this Manual The Service Manual for this unit can be supplied in the following languages Designation Part No Service Manual Perfusor compact German 8713 9111 The com...

Страница 3: ...cessories Page 1 3 Software Approved Software Versions Page 2 1 Version Display during Switch On Test Page 2 1 Extended Version Display during Switch On Test Page 2 1 Error Messages and Alarms Page 2...

Страница 4: ...pection Page 5 2 Electrical Safety Page 5 5 Maintenance Page 6 1 Technical Safety Check TSC Page 7 1 Procedural Instructions on the TSC Visual Inspection Page 8 1 Functional Inspection Page 8 2 Pressu...

Страница 5: ...the periods stated 6 MP BetreibV B Braun also recommends training on the Technical Safety Checks or to perform at least the steps indicated in the current version of the manual as the TSC requires tha...

Страница 6: ...n service the technician has participated in a technical training course for the specific B Braun unit Quality Management B Braun is certified in accordance with DIN EN ISO 9001 and ISO 13485 This cer...

Страница 7: ...e it is recommended that personnel wear cotton clothing and electrically conductive shoes to prevent electrostatic charge Spare Parts and Test Equipment Only use original spare parts from the manufact...

Страница 8: ...e figure When the Service Manual is stored as pdf file these references are displayed green Click with the mouse button on a reference to jump to the corresponding source Menu commands are described a...

Страница 9: ...ct your local B Braun subsidary International Technicians Intercompany Irene Marchel Fax 49 0 5661 75 38 57 e mail irene marchel bbraun com Service Hotline Martin Heusner Phone 49 0 56 61 71 35 25 Fax...

Страница 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes 2 1...

Страница 11: ...e keyboard The syringes are changed semi automatically The function process and monitoring is microprocessor controlled The Perfusor compact has a long service life and is easy to service due to its m...

Страница 12: ...erface 2 E Module as operating and control unit 3 Drive unit consisting of drive board with the complete sensor technology light barriers for syringe pre and end alarm syringe size recognition and mot...

Страница 13: ...1 Perfusor compact 2 1 gb 1 3 Accessories Designation Ord No Unit connecting lead 200 240 V 3450 2718 Unit connecting lead 100 240 V 3450 5423 Pole clamp universal clamp rotating 3450 9054 Battery pac...

Страница 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes 2 1...

Страница 15: ...after the other on screen 888 8 111 1 222 2 555 5 AA Reference to the instructions for use hard and soft ware group 3 The Perfusor compact switches over to normal operation Extended Version Display d...

Страница 16: ...CMP condition 3 Rate of FP and CMP different 4 Different function mode 5 Different rate of delivery 6 Different target volume 7 Different step volume low 8 Different motor steps 19 State motor state 2...

Страница 17: ...C H8 85 Bw difference between NEC H8 86 Md difference between NEC H8 90 Syringe state in Oper_Syr 91 Set syringe type 92 Consistency error 93 Difference between setting and display 94 Timer synchroniz...

Страница 18: ...150 Different software versions 151 Double CRC error 153 Different states 154 Different rates 155 Different F mode 156 Different mode values 157 Different alarm recognition 158 Different alarm clearan...

Страница 19: ...Plc_Down 208 No sync at Plc_On 209 CMP FP timer end sync error 220 Different phases busy 221 Different phases idle 222 Motor on at reverse steps 223 Too many pending steps 224 Motor current error 225...

Страница 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes 2 1...

Страница 21: ...ustrations of each user interface and where the description contains differences Introduction The Service Program runs in a user friendly way on a PC All functions are easy to operate in the pulldown...

Страница 22: ...s installed and the PC is connected to the Perfusor compact the following functions can be executed Drive calibration Reading loading pump data Displaying service values Displaying and changing parame...

Страница 23: ...saved in this directory if no other directory is specified The directory name can be changed without any problems Note The system configuration of the PC is not changed when the Service Program was in...

Страница 24: ...File Connect 2 Menu item EEPROM Read 3 Menu item File Save 4 Version 5 5004 Select menu item Modi Modification Data and menu item Syringe Syringe selection or Syringe types Note down parameters prior...

Страница 25: ...Program contains the Default dat file with the factory settings of the unit These values can be adjusted via the Syringe or Modes menu if required Max delivery rate basal rate 99 9 ml h Bolus rate 80...

Страница 26: ...accept any syringe after a service was carried out Is syringe selection set to free type but free type was not loaded Then Set syringe selection to table OPS or load corresponding syringe Problems und...

Страница 27: ...on a freely selectable connected printer 3 Save F8 Saves the unit data e g on the hard disk Enter the user number 0 upon query 4 Configuration For selecting language and interface Also to select the...

Страница 28: ...e Program and overwrites all previous settings Therefore settings which can be changed should be read out and noted down see Checklist after Operation of the Service Program p 3 18 Recalibrate the uni...

Страница 29: ...s determined by pressure calibration 4 Overload check see Overload Check p 3 17 The dynamic pressure test is used to determine whether the unit was damaged after having been dropped due to a shock or...

Страница 30: ...type Allows you to select all syringe types which are saved and can be loaded in the Perfusor compact OPS Defines that only OPS 50 ml and OPS 20 ml syringes can be used 2 Free type 20ml and Free type...

Страница 31: ...necessary adapt the syringe selection e g when the selection was set to Free type but all free types were deleted Note Adapt the syringe table after all modifications have been carried out see Syringe...

Страница 32: ...be carried out if the bolus rate was changed Constants Menu 1 Service interval Reads and resets the service interval timer When the time set has elapsed a service interval alarm is triggered when the...

Страница 33: ...Acknowledge by starting again Now the unit delivers the maximum basal rate Bolus Rate if not blocked 1 Press the F1 key bolus rate The value displayed must correspond to the bolus rate set 2 Start bo...

Страница 34: ...e down the current code for 20 ml and 50 ml syringes before starting the test Test for table or internal setting 1 Open holder then press keys 7 C X X F and 7 C Y Y F The syringe is accepted if an ala...

Страница 35: ...of the 20 ml free type F and keys 7 C number of the 50 ml free type F syringe is accepted if an alarm is triggered then the syringe selection setting is not correct Test for OPS setting 6 Open holder...

Страница 36: ...ommunication via menu item File Connect F1 2 After activities on the drive Transfer data via menu item EEPROM Read F3 from the unit to PC 3 After replacement of the E Module Call default data from the...

Страница 37: ...Check unit according to the procedure instructions for inspection see Procedural Instructions for Inspection after Modifications via the Service Program p 3 13 Overload Check 1 Connect 50 ml OPS syrin...

Страница 38: ...atic with off alarm at near empty alarm with power on test dynamic static with off alarm at near empty alarm with power on test Modes Sound of alarms Modes Sound of near empty alarms Syringes Syringe...

Страница 39: ...022 is displayed upon startup Press the ON OFF button to delete the alarm 022 until the alarm symbol is no longer displayed If the alarm 105 is triggered afterwards switch the unit off Before startup...

Страница 40: ...t for quick reference guide 5 screw split rivet for battery compartment cover 5 blind plug for syringe holder 5 countersunk screw M 3x10 5 flat head screw M 3x5 10 flat head screw M 3x6 5 board holder...

Страница 41: ...E Module when disconnecting 4 Dismount both housing halves Always check A Module before replacing the board Other modules can only be exchanged without danger of consequential damage if there is no ov...

Страница 42: ...ecks after repair Procedural instructions see Procedural Instructions for Inspection after Modifications via the Service Program p 3 13 A calibration in the Service Program is to be carried out if a n...

Страница 43: ...1329 Spanish 3891 1337 Dutch 3891 1345 Italian 3891 1353 Portuguese 3891 1361 Norwegian 3891 1370 Finnish 3891 1388 Swedish 3891 1396 Czech 3891 1400 Turkish 3891 1418 Polish 3891 1426 Danish 3891 143...

Страница 44: ...holder with pin punch 3 Remove screw 4 Pull off holder 5 Push spacer washer 0 8x4x3 2 on the shaft if the syringe is not recognized 6 Insert new syringe holder 7 Fit new screw not the old one and safe...

Страница 45: ...the battery compartment cover does not get jammed Check for tight fit The battery compartment cover is also the holder plate for the pole fixation 4 6 Snap in Clip Designation Ord No Snap in clip and...

Страница 46: ...and press MFC socket inwards 3 Pull off the N Module connector slightly pull out the A Module 4 Pull off connector on the E Module 5 Replace A Module and check snap in hook on the board 6 Assembly is...

Страница 47: ...y compartment see Battery Compartment Cover p 4 7 2 Remove batteries 3 Connect unit to mains power and switch unit on 4 Disconnect from mains power for a short moment when the switch on test is finish...

Страница 48: ...nnection cable 7 Insert new E Module at the side into the guide and position behind the holder Caution Do not damage the components 8 Push board in the guide to the right and insert a new board holder...

Страница 49: ...embly is performed in reverse order Note Lay two wire cable with mains power connector behind bearing Connect mains power connector correctly to the A Module please see figure Do not squeeze the cable...

Страница 50: ...lock 3450 9100 WARNING The drive consists of safety relevant parts FUNCTIONAL SAFETY JEOPARDIZED IF PARTIALLY REPLACED Always replace entire drive Exchange 1 Open unit see Open unit p 4 3 2 Move drive...

Страница 51: ...able 10 Calibrate in Service Program see Unit Calibration p 3 16 4 14 Axial Positioner Designation Ord No Axial positioner 3450 6659 Exchange 1 Open unit see Open unit p 4 3 2 Move drive arm to middle...

Страница 52: ...le and slide until stopper of the aluminum profile from the motor side CAUTION Cable layout according to figure 7 Press board against stopper when screwing down Tighten screws hand tight 8 Fix satelli...

Страница 53: ...p 3 16 4 16 Drive Head Designation Ord No Drive head complete 3450 1720 Toggle 3450 1711 Exchange 1 Open toggle pull out drive head and close toggle again 2 Pierce tamper proof caps in drive head with...

Страница 54: ...lug connector 8 Push drive head housing toward the unit 9 Pull jaws off the drive tube 10 Pull drive head housing together with clamp off the drive tube 11 Assembly is performed in reverse order Note...

Страница 55: ...too hot as otherwise the housing is damaged b Clean adhesive position on new housing and stick type plate New type plates can only be ordered as spare parts if the old type plates are returned to B Br...

Страница 56: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb 2 1 For your notes...

Страница 57: ...ead Syringe table quick reference guide Membrane keyboard Battery compartment cover battery compartment and contacts Unit feet MFC connector Holder for pole fixation side snap in mechanism Mains power...

Страница 58: ...Inspection Mechanical Inspection 1 Check function of the holder for pole fixation 2 Check stacking function of the unit with respect to other units 3 Check function of the syringe holder with syringe...

Страница 59: ...2 3 ml h key sequence 1 2 3 The pump delivers The delivery rate set must be displayed 5 Open lock Alarm by buzzer and positive locking sensor alarm Drive stops Lock must not engage in upper position 6...

Страница 60: ...deliver at 96 0 ml h and switches off when the specified pressure is reached 9 Enter value in check list WARNING The calibration gauge can release suddenly when unlocked manually DANGER OF INJURY Do n...

Страница 61: ...20 ml OPS syringe to 6 ml and insert syringe 6 Confirm syringe type 20 with F button 7 Press START button Pump delivers and triggers pre alarm at a volume of 5 ml 8 Bolus press buttons F and 1 simulta...

Страница 62: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb 2 1 For your notes...

Страница 63: ...b 6 1 6 6 Maintenance The unit is maintenance free A Technical Safety Check TSC is to be carried out every 24 months to check the operational capability of the Perfusor compact see Technical Safety Ch...

Страница 64: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes 2 1...

Страница 65: ...de snap in mechanism 2 Functional Inspection Switch on unit Compare with instructions for use LCD self test and audible alarm Compare set delivery rate and value displayed Check switching capability o...

Страница 66: ...ort circuited mains inlet and plus pole in battery compartment top right 6 Accessories Enter MFC battery etc 7 Optional Rate limitation ml h Bolus rate limitation ml h Infusion line used for Technical...

Страница 67: ...function with OPS 50 ml syringe Syringe holder axial positioner drive head clamp and push button sensor Membrane keyboard Check adhesion cleanliness and fit Battery compartment cover and battery cont...

Страница 68: ...display if a syringe was not inserted 5 Insert current step gauge in the OPS 50 ml slot and close syringe holder The lock must engage automatically and the syringe symbol in the LC display must not fl...

Страница 69: ...espond to the TSC values 2 Pump continues to deliver at 96 0 ml h and switches off when the specified pressure is reached 3 Enter value in check list 4 Set pressure stage 2 key sequence F 3 2 F START...

Страница 70: ...Remove potentiometer calibration gauge c Insert max 50 ml potentiometer calibration gauges The 50 ml syringe symbol is displayed d Remove potentiometer calibration gauge Electrical Safety 1 Measure t...

Страница 71: ...Procedural Instructions on the TSC 8 Perfusor compact 2 1 gb 8 5 Optional Enter rate limitation and bolus rate limitation in TSC 2 1...

Страница 72: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb 2 1 For your notes...

Страница 73: ...1703 Calibration gauge 6 6 80 N 0770 1535 Current step gauge 10 130 N 0770 1616 for TSC as option Potentiometer calibration gauge 20 ml 0770 1543 Potentiometer calibration gauge max 50 ml 0770 1624 MF...

Страница 74: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes 2 1...

Страница 75: ...3891 1396 Czech 3891 1400 Turkish 3891 1418 Polish 3891 1426 Danish 3891 1434 Syringe holder complete 3450 6608 with screw and cap Unit feet 3450 6640 Battery compartment cover 3450 6632 Snap in clip...

Страница 76: ...t pin lock 3450 9100 Axial positioner 3450 6659 Drive board 3450 6691 with main PCB and satellite boards for syringe size recognition and recognition of direction of rotation Drive head complete 3450...

Страница 77: ...41 Safety hook 3450 8368 Plate 2 pcs 3450 2610 Connection cap D12 4 5 pcs 3477 4149 Bellows 5 pcs 3477 3274 Pressure spring 5 pcs 3477 4165 Universal Clamp Universal clamp complete not available any m...

Страница 78: ...Spare Parts List 10 10 4 Perfusor compact 2 1 gb For your notes 2 1...

Страница 79: ...006 Changes in this version Checking the pressure cut off during Bolus after repair Version 3 0 This Service Manual was approved by B Braun on 20 08 2009 Changes in this version New Windows based soft...

Страница 80: ...Revision Documentation 11 11 2 Perfusor compact 2 1 gb 2 1...

Страница 81: ...e 3 5 Version 3 0 Page 3 6 Version 3 0 Page 3 7 Version 3 0 Page 3 8 Version 3 0 Page 3 9 Version 3 0 Page 3 10 Version 3 0 Page 3 11 Version 3 0 Page 3 12 Version 3 0 Page 3 13 Version 2 1 Page 3 14...

Страница 82: ...ersion 2 1 Page 8 3 Version 3 0 Page 8 4 Version 3 0 Page 8 5 Version 2 1 Page 8 6 Version 2 1 Chapter 9 Test Equipment and Special Tools Page 9 1 Version 2 1 Page 9 2 Version 2 1 Chapter 10 Spare Par...

Страница 83: ...odule 4 10 End alarm 5 5 Error messages 2 2 F Function 1 2 Functional Inspection 5 2 8 2 H Housing bottom part 4 17 Housing upper part 4 11 L List of abbreviations 0 8 LS clip 4 9 N N Module 4 11 O Op...

Страница 84: ...2 Perfusor compact 2 1 gb Syringe table 4 5 Syringes 8 4 T Technical Safety Checks 0 5 7 1 Test equipment 9 1 Trouble shooting 3 5 TSC 0 5 7 1 U Unit feet 4 6 V Version display 2 1 Visual Inspection...

Страница 85: ......

Страница 86: ...B Braun Melsungen AG Carl Braun Stra e 1 34209 Melsungen Germany Tel 49 0 5661 71 0 Fax 49 0 5661 71 4567 www bbraun de No 8713 9112 Version 3 1 Effective 14 03 2012...

Отзывы: