![Braun perfusor compact Скачать руководство пользователя страница 65](http://html1.mh-extra.com/html/braun/perfusor-compact/perfusor-compact_service-manual_2804981065.webp)
Index 01
(Master - to be added to the documentation)
Perfusor® compact, 2.1 gb
7 - 1
Sheet 1 of 2
M659 01 00 06F04 01 / 3891 1128
7 -
Technical Safety Check TSC
Checklist for Technical Safety Checks – Every 24 Months
Unit: Perfusor compact infusion syringe pump
Manufacturer: B. Braun Melsungen AG
User
Observe the Service Manual and the instructions for use. All measured values are to be documented.
Accessories used should be included in testing. Make exclusive use of calibrated measuring
equipment.
Item No.
Unit No.
Year of Procurement
1. Visual Inspection
❒
Unit clean, complete,
undamaged
❒
Syringe fastening:
Syringe holder, axial
positioner, drive head,
clamp, push-button sensor
❒
Membrane keyboard
❒
Battery compartment cover
and battery contacts
❒
Unit feet
❒
Mains power lead and
connector
❒
MFC lead and connector
❒
Holder for pole fixation, side
snap-in mechanism
2. Functional Inspection
Switch on unit.
❒
Compare with instructions
for use: LCD self-test and
audible alarm
❒
Compare: set delivery rate
and value displayed
❒
Check switching capability
of staff call (accessories)
❒
Switch on unit in battery
mode and check self-test
Note
If "Battery discharged" is
displayed:
Charge battery or replace
batteries and repeat test.
❒
Compare status display
000"A" or 000"b" with
battery cells used.
❒
Check push-button sensor
alarm
❒
Check positive locking
sensor alarm
3. Pressure cut-off
(alternatively with manometer or check
gauge)
With manometer and 50 ml
syringe:
❒
Pressure stage 1 <0.6 bar
❒
Pressure stage 2 <0.9 bar
❒
Pressure stage 3 <1.2 bar
With check gauge current step
Order - No. 0770 1616:
❒
Pressure stage 1 <40 N
❒
Pressure stage 2 <59 N
❒
Pressure stage 3 <75 N
CAUTION
Danger of injury: Remove check
gauge only when released.
4. Syringes
Syringe selection
❒
OPS
❒
internal
❒
EEprom
Syringe table readable
❒
Yes
❒
No
Syringe recognition
Manufacturer (code) used
❒
20 ml
❒
50 ml
❒
_____
❒
_____
(Part 1 of 2)
7 - 2
2.1
Содержание perfusor compact
Страница 1: ...Perfusor compact Service Manual 0 10 3 1 0 Version 3 1 English...
Страница 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes 2 1...
Страница 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes 2 1...
Страница 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes 2 1...
Страница 56: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb 2 1 For your notes...
Страница 62: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb 2 1 For your notes...
Страница 64: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes 2 1...
Страница 72: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb 2 1 For your notes...
Страница 74: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes 2 1...
Страница 78: ...Spare Parts List 10 10 4 Perfusor compact 2 1 gb For your notes 2 1...
Страница 80: ...Revision Documentation 11 11 2 Perfusor compact 2 1 gb 2 1...
Страница 85: ......