Checks after Repair
5
5 - 4
Perfusor® compact, 3.0 gb
Pressure Cut-Off
CAUTION
The limit values of the Checklist for Checks after Repair do not
correspond to the TSC values.
1. Set pressure stage 1 (key sequence F 3 1 F START).
2. Pump continues to deliver at 96.0 ml/h and switches off when
the specified pressure is reached.
3. Enter value in check list.
4. Set pressure stage 2 (key sequence F 3 2 F START).
5. Pump continues to deliver at 96.0 ml/h and switches off when
the specified pressure is reached.
6. Enter value in check list.
7. Set pressure stage 3 (key sequence F 3 3 F START).
8. Pump continues to deliver at 96.0 ml/h and switches off when
the specified pressure is reached.
9. Enter value in check list.
WARNING
!
The calibration gauge can release suddenly when unlocked
manually.
DANGER OF INJURY!
-
Do no open toggle.
10. Release calibration gauge (key sequence F 3 0) with service
connector.
11. Re-start delivery.
12. Press "F" and "1" keys simultaneously and keep them pressed.
13. Pump continues to deliver at set bolus rate until pressure
alarm ist triggered and switches off when the specified
pressure is reached.
14. Enter value in check list.
15. Release calibration gauge (key sequence F 3 0). Do not
remove calibration gauge until it is fully released.
Syringe Recognition
1. Insert 20 ml potentiometer calibration gauges.
The 20 ml syringe symbol is displayed.
2. Remove potentiometer calibration gauge.
3.0
Содержание perfusor compact
Страница 1: ...Perfusor compact Service Manual 0 10 3 1 0 Version 3 1 English...
Страница 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes 2 1...
Страница 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes 2 1...
Страница 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes 2 1...
Страница 56: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb 2 1 For your notes...
Страница 62: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb 2 1 For your notes...
Страница 64: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes 2 1...
Страница 72: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb 2 1 For your notes...
Страница 74: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes 2 1...
Страница 78: ...Spare Parts List 10 10 4 Perfusor compact 2 1 gb For your notes 2 1...
Страница 80: ...Revision Documentation 11 11 2 Perfusor compact 2 1 gb 2 1...
Страница 85: ......