Operation 2016-03-07
Built-in Electrode Replacement
• Always use the guide bur and make sure it will not come out. If the guide bur is not properly fix in place, the internal contact could be bent, and then the instru-
ment might not be able to make accurate measurements or else it might malfunction.
• Do not run the motor with the guide bur inserted; this could damage the instrument.
Take out the screw and then take out the built-in electrode.
Hold the push button and slide the guide bur straight in as
shown in the illustration. Then rotate it left and right.
Put a little Ethanol for Disinfection (Ethanol 70 to 80 vol%)
on a brush and clean the rotor axle with it.
Slide the built-in electrode onto the guide bur and line up
the screw holes.
Blow air on the electrode to remove any remaining mois-
ture.
Rotor Axle
Slowly turn the screw and make sure the built-in electrode
goes into the head properly.