14
Operation 2016-03-07
Additional Functions:
Apical Action Function (APICAL ACT.**), Apical Reverse or Stop (APICAL ACT.**)
Auto Start or Stop (AUTO STT/STP**)
1. Select a memory number for the standby display; press the Plus or Minus switch.
2. Hold down the Select switch for at least 1 second to show the displays for additional opera-
tion settings.
3. Press the Select switch to go from one display to the next.
4. Change the setting; press the Plus or Minus switch.
* The display will go back to the standby display if 5 seconds (initial setting) elapses without
a switch being pressed.
Standby Display
Apical Action Function
**
:
When this is turned ON, Apical Reverse or Stop function be activated.
Apical Reverse or Stop
**
:
The file will reverse or stop when the file tip reaches the Flash Bar.
When Apical Action Function is turned OFF, this display will be skipped.
Auto Start & Stop
**
:
When this is turned ON, the file starts rotating when it is inserted and stops when it is taken
out of the canal.
**
These functions are available only when connected to the EndoSync A.I.
Memory Settings: Additional Operation Settings (OTR mode)
Apical Action Funciton
Apical Reverse or Stop
Auto Start & Stop