BRASSELER USA Carbide burs Скачать руководство пользователя страница 1

Brasseler USA CARBIDE Burs Instructions for Use 

 

 

 

P 1 of 3 

 

IFU-0006 REV A 

 

 

EDCO 11.0098

 

BRASSELER USA CARBIDE Burs 

BRASSELER USA Carbide Burs are available in Sterile and Non-Sterile models under various trade names with numerous 
head diameters and working lengths. The devices are reusable and are sterilized using steam sterilization in a gravity or 
prevacuum cycle.   

Description 

The BRASSELER USA CARBIDE Dental Bur family includes Clinical Carbides and Laboratory Carbides. A CARBIDE dental 
bur is a rotary cutting device made of stainless steel, the working end of which is made from tungsten carbide, and 
which is designed to fit into a dental handpiece. Alternatively, the device may be constructed wholly from tungsten 
carbide. BRASSELER USA CARBIDE Burs are reusable devices.  

Intended Use 

BRASSELER USA CARBIDE Burs fit into a dental handpiece, which provides the rotation, allowing the user to cut hard 
structures in the mouth, e.g., teeth or bone. BRASSELER USA CARBIDE Burs can also be used to cut hard metals, plastics, 
porcelains and similar materials.  

Warnings and Precautions 

 

Warning

: Attention should be paid to the speed of work (RPM)

 

 

o

 

Always refer to the product packaging for the Maximum RPM

 

o

 

Operating a bur with too high of an RPM may generate undesirable heat

 

o

 

Operating a bur with too high of an RPM may cause the bur to fail

 

 

 

CARBIDE Burs must be thoroughly cleaned and steam sterilized prior to the first use and each subsequent reuse  

 

CARBIDE Burs labeled “Sterile” require no further action prior to first use but must be thoroughly cleaned and 
steam sterilized prior to each subsequent reuse   

o

 

If the packaging for “Sterile” labeled devices is opened or damaged, the device must be thoroughly 
cleaned and steam sterilized prior to use and each subsequent reuse  

 

Do not use chemical or dry heat to sterilize BRASSELER USA CARBIDE Burs, as these processes have not been 
validated for use 

 

Do not use worn-out or dull burs 

 

Do not apply excessive pressure on the bur as this could cause undesirable heat or may cause the bur to fail 

 

Avoid removing the bur at too sharp an angle to avoid leverage and breakage 

 

Carefully read package labels to ensure use of the appropriate device  

 

Ensure the bur is fully seated and securely gripped in the handpiece collet prior to use  

 

Move the bur continuously when in use to avoid localized heating and/or damage to the bur 

 

Maintain handpieces in good working condition to ensure maximum effectiveness of the device 

 

Use a rubber dental dam while using CARBIDE Burs to avoid possible aspiration or swallowing 

 

Always wear gloves when handling contaminated instruments 

 

Eye protection must be worn to protect against eject particles 

 

Surgical masks must be worn to avoid inhalation of any aerosol or dust generated 

 

BRASSELER USA Bur Blocks used to hold the devices for storage and steam sterilization are not intended to 
maintain sterility of the device 
 