6-4
100-214-239 Rev. A
Operation
The following chart summarizes the System Mode options, including the use of Limits:
NOTE
i
You must have a Temperature Probe attached to the unit to use the Temperature Mode, the
Pulse/Pause Mode, or to use Temperature as a Limit.
System States
Availability of the Digital Sonifier’s functions depends on the system state. The ‘normal’ state of the system for
most users is the Run or Ready mode. Some remote terminal commands are noted where appropriate (com-
mands have the format of an ‘!’ followed by upper-case letter or letters, i.e. !TS for the Test command). These
commands are also summarized in the Remote Terminal section found on page 6-35. The possible system states
include:
• Power-Up: The system enters the Power-Up state when the power switch is turned on. The fan starts to
run, and the system then enters the Self-Diagnostics state.
• Self-Diagnostics: In this state, the system performs hardware and software checks on itself. The front
panel display shows several diagnostic displays (indicating which step in the self-diagnostics it is in) and
the front panel LEDs will illuminate as a visible test. After it successfully completes this system check, it
enters the Ready state.
• Ready: In the Ready state, the system is waiting to begin an experiment, and it can accept parameter
settings that you enter through the keypad or the remote computer or terminal. The front panel display
Time
setting
Pulse On and
Pulse Off
settings
Max. Temp
setting
Pulse Temp.
setting
Cycle Ends
upon...
Continuous
Mode
OFF
Both OFF
OFF
OFF
Stop button or
remote command
Timed Mode
Your Time
parameter
Both OFF
OFF or Limit
OFF
Time parameter
elapses
Pulse Mode
OFF or Limit
Pulse On and
Pulse Off
times
OFF or Limit
OFF
Stop button or
remote command
Temperature
Mode
OFF or Limit
Both OFF
maximum
Temperature
parameter
OFF
Maximum
Temperature
reached
Pulse/Pause
Mode
OFF or Limit
Optional.
The system
will recalculate
while running
the process
Max. Temp
set to upper
temperature
for process
Pulse Temp
set to lower
temperature
for process
Stop button or
remote command
Содержание Sonifier 250
Страница 4: ...100 214 239 Rev A For Your Notes...
Страница 5: ...100 214 239 Rev A i 5 Digital Sonifier...
Страница 13: ...1 8 100 214 239 Safety and Support For Your Notes...
Страница 23: ...2 10 100 214 239 Rev A Introduction to the Digital Sonifier For Your Notes...
Страница 25: ...3 2 100 214 239 Rev A Delivery and Handling For Your Notes...
Страница 101: ...7 10 100 214 239 Rev A Maintenance For Your Notes...
Страница 121: ...C 4 100 214 239 Rev A Accessories Parts List For Your Notes...
Страница 125: ...D 4 100 214 239 Rev A Wyse 55 Terminal Settings For Your Notes...