![Brainlab CRANIAL/ENT OPTICAL TRACKING Скачать руководство пользователя страница 60](http://html.mh-extra.com/html/brainlab/cranial-ent-optical-tracking/cranial-ent-optical-tracking_instrument-user-manual_2800942060.webp)
5.3
Vario Reference Arm (Mayfield Headholder)
Introduction
The
Vario Reference Arm (Mayfield Headholder)
(52001) creates a flexible and extended
attachment of the
Standard Cranial Reference Array
onto the Mayfield Modified Skull Clamp.
Compatible Reference Arrays
Name
Article No.
Standard Cranial Reference Array with 4 Marker Spheres
41725
How to Attach the Array to the Reference Arm
①
②
③
④
⑤
Figure 21
Step
1.
Attach the unsterile
Standard Cranial Reference Array
⑤
to the array connector
③
, se-
curing it with the fixation nut
④
.
2.
To position the array, turn the central tension handle
②
counterclockwise
to loosen it.
3.
Adjust the arm
①
to the desired position and securely re-tighten the handle by turning it
clockwise
.
Loosening the Arm
Warning
Do not grip the Vario Reference Arm (Mayfield Headholder) at the ball joints while the
central tension handle is unfastened as you could injure your fingers.
Warning
Never loosen the central tension handle without gripping the articulated arm with your
other hand. The Vario Reference Arm could fall on the patient.
Warning
Loosen the mechanism by turning the central tension handle counterclockwise until it
stops and avoid using any additional force. Hold the articulated arm firmly with one hand
Vario Reference Arm (Mayfield Headholder)
60
Instrument User Guide Rev. 2.6 Cranial/ENT Optical Tracking
Содержание CRANIAL/ENT OPTICAL TRACKING
Страница 2: ......
Страница 20: ...Available Instruments and Classification 20 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Страница 28: ...Instrument Calibration Matrix 28 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Страница 52: ...Electromagnetic Compatibility 52 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Страница 86: ...Instrument Adapter Clamp Softgrip Size L XL 86 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Страница 94: ...Ultrasound Calibration 94 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Страница 102: ...Attaching the VarioGuide to the Head Holder 102 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Страница 110: ......
Страница 111: ......
Страница 112: ...brainlab com Art No 60921 34US 60921 34US ...