Dialflow Regulator 

Instructions for Use 

 

 

 

 

 

Содержание Dialflow Regulator

Страница 1: ...702 0084 14 October 2014 Dialflow Regulator Instructions for Use ...

Страница 2: ...ug and may be used in critical treatments Medical gases should only be used for medical purposes on the authority of a physician or authorised clinician and then strictly in accordance with their instructions Oxygen is not flammable however an oxygen enriched atmosphere will drastically increase the rate and severity of combustion Oil and or grease in the presence of an oxygen enriched atmosphere ...

Страница 3: ...all stand or cart Do not use or store oxygen equipment near excessive heat 50 C or 125 F or open flame Only appropriately trained personnel working in controlled conditions must perform disassembly assembly and testing of regulators and flowmeters Dialflow Regulators are not MRI compatible Flowmeters may be affected if the cylinder pressure is other than stated in the specifications The threaded b...

Страница 4: ...nd these may be different from one country to another ensure that this regulator is the correct connector for the cylinder to be used Once connected to an oxygen cylinder and with the cylinder valve opened the user may select the flow rate prescribed by rotating the gas flow selection dial until the correct flow rate is seen in the viewing window The rate at which gas flows from the regulator is c...

Страница 5: ...4 4 After Use Upon completion of the therapy turn the flow selection dial to zero close the cylinder valve and disconnect the outlet hose from the regulator and the patient To remove the regulator from the cylinder first ensure the cylinder valve is closed Bleed off any residual gas pressure in the regulator by turning the flow selection dial to any flow setting until flow ceases Remove the regula...

Страница 6: ... service in accordance with the instructions given in the Service Manual or return the device to a recognised BPR Medical Service Centre on or before the date shown 6 Specification Specification Value Flow Accuracy 15 of setting for flow rates above 1 L min 20 10 of settings of 1 L min and below Input Pressure Range 900 to 20 000 kPa 9 bar to 200 bar Flow Settings l min Range A 0 0 02 0 03 0 05 0 ...

Страница 7: ...ation to be supplied General requirements BS EN 13544 2 A1 Respiratory therapy equipment Tubing and connectors BS 5682 Specification for probes quick connectors for use with medical gas pipeline systems 7 Troubleshooting Fault Possible Cause Solution No gas flow Gas cylinder turned off Check gas supply Dial Flow Regulator not connected properly Check gas specific probe is correctly connected Gas c...

Страница 8: ...cing 212 0012 Pin Index Yoke Seal 212 0048 Bullnose seal Benelux 303 0030 Cylinder Gauge Boot 303 0060 Swivel Cylinder Gauge 0 3mm Restriction 9 Distributor Details Australia BOC Healthcare Customer Service Centre 10 Julius Avenue North Ryde Sydney NSW 2113 T 61 0 2 8874 4400 F 61 0 2 9886 9000 www boc com au New Zealand BOC Healthcare Customer Service Centre 970 992 Great South Road Penrose 1061 ...