6 Operation
24
BOWA-IFU-MN031-554-ERGO315R-S3-EN-20210219
6.
Operation
6.1.
Before use
The vessel sealing instrument is assembled (see section 5,
page 14) and reprocessed (see section 8, page 32).
WARNING
Risk of patient injury!
Use only approved BOWA ARC generators with vessel
sealing capability (see section 9, page 45).
Select the settings of the HF device according to the
requirements of the intervention.
Use only suitable devices and accessories as described
in the key.
Use only intact and sterilised devices.
Risk of patient injuries from burns or explosion!
Avoid contact with flammable gases and liquids, such
as skin cleansers, disinfectants and anaesthetic gases.
Avoid direct skin contact with HF cables.
Skin-to-skin contact (e.g. between the patient's arms
and body) should be avoided, e.g. by inserting dry
gauze.
1.
Switch on the HF device and connect the HF cable to the HF device.
2.
Set the required settings on the HF device.
3.
Before each use of the vessel sealing instrument, carry out a
thorough visual and functional check (see section 6.2, page 25