5 Operation
48
Operating Manual ARC 400
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
5.6.5. Resection
This mode is used in gynecology and urology. ARC control generates the cutting
effect with simultaneously minimized output power. ARC control facilitates direct
cutting and prevents electrode adhesion.
Use non-conductive irrigation fluids.
Application areas
Hysteroscopy, transurethral prostate resection (TUR-P), surgical treatment of
bladder tumors (TUR-B), vaporization of prostate tissue (TUR-VAP).
Suitable instruments
Resectoscope (monopolar)
Resection sling
Rollerblade electrode
5.6.6. MetraLOOP
This mode is used in gynecology for laparoscopic hysterectomy. Removal of the
uterus can be achieved by applying monopolar cutting current and pulling on the
sling at the same time.
Application areas
Gynecology; laparoscopic hysterectomy
Suitable instruments