11 Technical specifications
900-100_IFU-V1.0_11587-S0-20141209-EN
ARC 100 Operating Manual
39
11.
Technical specifications
11.1.
Technical data for ARC 100
Insulation type / Classification
EMC
IEC 60601-1-2: 2007
Level of protection provided by the housing
IP 21
Protection class according to EN 60601-1
I
Type of application component as specified in EN
60601-1
CF
Compliance with standards
IEC 60601-1: 2005, +
Cor. 1:2006 + Cor. 2:2007 + A1:2013
IEC 60601-1-2: 2007,
IEC 60601-2-2: 2009,
IEC 60601-1-6:2010
IEC 60529:1989 + A1:1999
IEC 62304:2006
IEC 62366:2007
ISO 14971: 2007,
ISO 13485: 2003 + Cor.1: 2009
Classification according to EC Directive 93/42/EEC
IIb
Power connection
Power consumption in standby mode (100 V)
14 VA
Power consumption in standby mode (230 V)
30 VA
Line frequency
50/60 Hz
Maximum power consumption with 100 W HF output
power
160 VA
Connection for potential equalisation line
Yes
Voltage range 100 - 260 V
Input voltage range
100 V to 260 V
Current consumption in standby mode (100 V)
140 mA
Current consumption in standby mode (230 V)
130 mA
Current consumption at maximum HF output (100 V)
1.6 A
Current consumption at maximum HF output (230 V)
0.7 A
Mains fuses
2 x T3,15 A 250V
Dimensions and weight
External dimensions: width x height x depth (mm)
280 x 114 x 310
Weight
Approx. 5.6 kg