2 Safety
12
ARC 100 Operating Manual
900-100_IFU-V1.0_11587-S0-20141209-EN
2.2.
General safety instructions
Ensure that no electronic devices that are subject to interference
from electromagnetic fields are set up in the vicinity of the HF
device.
Please comply with the instructions on electromagnetic
compatibility (EMC); see Section EMC, page 47.
Always connect the HF device to a mains power system with a
protective earth lead in order to prevent electric shock.
Additional devices that are connected to electrical medical devices must
demonstrably satisfy the relevant IEC or ISO standards (e.g. IEC 60950
for data processing devices). Furthermore, all configurations must
comply with the standardised requirements for medical systems (see
IEC 60601-1-1 or Section 16 of the 3rd edition of IEC 60601-1, as
relevant). Anyone who connects additional devices to medical electrical
devices is perforce a system configurator and therefore responsible for
meeting standardised system requirements. Please note that local laws
prevail over the aforementioned standard requirements. For further
advice, please contact your local specialist retailer or our technical
service; see Section Technical service, page 38.
To protect personnel, BOWA recommends the use of a smoke evacuator
to extract electrosurgical smoke, e.g. BOWA SHE SHA.