Страница 2: ...1 User s Guide Bovie Smoke Shark Surgical Smoke Evacuator USER S GUIDE...

Страница 3: ...tructions 7 Filter Removal Instructions 8 Setup and Operation 8 Specifications for Smoke Evacuator System 9 Maintenance 10 General Maintenance Information 10 Cleaning 10 Periodic Inspection 10 Trouble...

Страница 4: ...acuator system preferentially removes toxic organic gases rather than water vapor and provides optimal odor removal The third stage filtration is a VLSI Very Large Scale Integrated grade ULPA Ultra Lo...

Страница 5: ...water 5 The smoke evacuator system has a classification under the UL2601 1 Clause 5 4 according to the method of sterilization or disinfection recommended by Bovie Medical Corporation Unplug the unit...

Страница 6: ...nit This device is not intended for evacuation of fluid If fluid is expected to be aspirated to the 18 hour filter fluid collection devices must be installed with the vacuum hose assembly Failure to i...

Страница 7: ...rane panel To power up the machine place the power switch on the side of the machine in the ON position and press the ON OFF switch on the mem brane panel until the suction control lights illuminate P...

Страница 8: ...is in standby suction motor off for greater than six 6 hours or the main power is switched off a new filter is required to activate the smoke evacuator and make it operational Fuses Side Panel Two 10...

Страница 9: ...le on side of machine 4 Ensure that the evacuation tubing is fully seated in the inlet of the filter 5 Turn on unit main power by placing the power switch on the side panel of the machine in the ON po...

Страница 10: ...nute 35mm NA 32mm 1671 22mm 1189 6mm 136 Variable Flow Control Yes Motor hp kW 1 4 0 75 Static Suction Inches H20 mbar 87 1 217 Hoses tubing 1 4 7 8 1 1 4 6mm 22mm 32mm Noise level dBA at min power 50...

Страница 11: ...g cleanliness and absence of damage to the filter inlet Obvious external or internal damage to the system Troubleshooting the System 10 Bovie Medical Corporation PROBLEM 2 Smoke Evacuation System does...

Страница 12: ...mmodate individual customer requests for shipping methods Bovie Medical Corporation reserves the right to decide shipping methods on prepaid orders Care is exercised in the checking and packaging of a...

Страница 13: ...rranties implied by any course of dealing or usage between parties are expressly excluded Consequential Damages Limits of Liability Bovie Medical Corporation shall not in any case whatsoever be liable...

Страница 14: ...ing read instructions Alternating current CAUTION Grounding reliability can only be achieved when the equipment is connected to a receptacle marked Hospital Grade or Hospital Only CAUTION Electric sho...

Страница 15: ...14 Bovie Medical Corporation...

Страница 16: ...2012 07 27...