Smoke Shark™ III

Surgical Smoke Evacuator

Operator’s Manual

www.symmetrysurgical.com                +1 615.964.5532 (International)                800.251.3000 (United States & Canada)                

Содержание Smoke Shark III

Страница 1: ...Smoke Shark III Surgical Smoke Evacuator Operator s Manual www symmetrysurgical com 1 615 964 5532 International 800 251 3000 United States Canada ...

Страница 2: ...Operational Information 1 4 Cautions and Warnings 2 0 OPERATING INSTRUCTIONS 9 2 1 Control Panel 2 2 Set up and Operation 2 3 Filter Instructions 2 4 Automatic Activation Device Instructions 2 5 Performance References 2 6 Electromagnetic Compatibility Information per IEC 60601 1 2 3 0 MAINTENANCE 16 3 1 General Maintenance Information 3 2 Cleaning 3 3 Periodic Inspection 3 4 Troubleshooting 4 0 CU...

Страница 3: ...ility can only be achieved when the equipment is connected to a receptacle marked Hospital Grade or Hospital Only CONMED Corporation 525 French Road Utica NY 13502 905215 Rev A EN Surgical Smoke Evacuator FR Évacuateur de fumée chirurgicale IT Evacuatore di fumi chirurgici DE Chirurgische Rauchabzugsvorrichtung ES Evacuador quirúrgico de humos EL Συσκευή εκκένωσης χειρουργικού καπνού PT Extrator d...

Страница 4: ...conductor Devices and equipment that are to benefit from this safety feature must have an appropriate three prong plug that is inserted into this outlet There are other possible arrangements for such outlets including the use of lateral grounding contacts that make contact with metallic strips on the side of the plug Also called by various other names including grounding outlet ground outlet groun...

Страница 5: ...ration media used to reduce the amount of activated carbon fines from migrating out of the filters 1 2 Inspection The Bovie Smoke Shark Surgical Smoke Evacuator was thoroughly tested and inspected prior to shipment Please inspect the Bovie Smoke Shark Surgical Smoke Evacuator before using to ensure all items have been received and that no damage occurred during transit If items are missing or dama...

Страница 6: ...acuator is void if any of the following warnings are disregarded Read this manual thoroughly and be familiar with its contents prior to using the Bovie Smoke Shark Surgical Smoke Evacuator Confirm operational set up of the Bovie Smoke Shark Surgical Smoke Evacuator prior to a surgical procedure Disconnect the Bovie Smoke Shark Surgical Smoke Evacuator from the grounded electrical outlet prior to i...

Страница 7: ...on could cause filter blockage electrical damage decreased efficiency and possible internal contamination The filter should be changed according to the Filter Life Indicator The filter should not be used for more than the filter life specified Failure to change the filter may result in decreased efficiency and possible internal contamination Care must be exercised in the installation and removal o...

Страница 8: ...L DANGER CAUTION RISK OF EXPLOSION IF USED IN THE PRESENCE OF FLAMMABLE ANESTHETICS CAUTION FEDERAL LAW USA RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN WARNING CONSULT INSTRUCTIONS DIRECTIONS FOR USE CAUTION MD MEDICAL DEVICE TYPE CF APPLIED PART LATEX NOT MADE WITH NATURAL RUBBER LATEX ALTERNATING CURRENT CE MARK PROTECTIVE EARTH GROUND EC REP AUTHORIZED REPRESENTATIVE IN THE ...

Страница 9: ...l provides a visual indication of the status of the filter life of the filter in use Install an unused filter into the Bovie Smoke Shark Surgical Smoke Evacuator as per the installation instructions section 2 3 When the Motor On LED is illuminated the Filter Life Indicator will light up the uppermost green LED indicating 100 Filter Life The indicator will progress through subsequent colors from gr...

Страница 10: ...ife remaining Failure to change the filter will affect the performance of the Bovie Smoke Shark Surgical Smoke Evacuator 2 3 Filter Instructions Bovie P N SF06 Configuration Wall Mounted Bovie SF06 Filter 4 Stage Filtration In One Casing Pre Filter ULPA Carbon Post Filter Filter s ULPA Particle Size µm 0 1 to 0 2 Microns At 99 999 Efficiency Filter Life 6 hours Automatic Factory Set Filter Sensor ...

Страница 11: ...e Smoke Shark III SE03 120 SE03 220 Maximum Flow Setting LPM U S Standard Hose I D 9 5 mm 3 8 SharkSkin Adapter 45 LPM Dimensions H x W x D inches 10 x 7 x 7 5 Dimensions H x W x D centimeters 25 4 x 17 8 x 19 1 Weight lbs 7 0 Weight kg 3 2 Noise Level dBA maximum 55 dBA 120 V 62 dBA 220 V Voltage Available 100 120 VAC 220 240 VAC Frequency auto sensed 50 60 Hz Remote Control Activation Yes option...

Страница 12: ...d in such an environment Emissions Test Compliance Electromagnetic Environment Guidance RF Emissions CISPR 11 Group 1 The Bovie Smoke Shark Surgical Smoke Evacuator uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF Emissions CISPR 11 Class A The Bovie Smoke Shark Surgical Smoke Evacua...

Страница 13: ...utput lines Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0...

Страница 14: ...nce in meters M IEC 61000 4 6 150 kHz to 80 MHz Field strength from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all sit...

Страница 15: ...l Smoke Evacuator as recommended below according to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 3 5 d P v1 80 kHz to 800 MHz 3 5 d P E1 800 kHz to 2 5 GHz 3 5 d P E1 0 01 0 12 0 12 0 23 0 1 0 34 0 34 0 74 1 1 7 1 7 2 3 10 3 7 3 7 7 4 100 11 7 11 7 23 3 For transmit...

Страница 16: ...ct both the Bovie Smoke Shark Surgical Smoke Evacuator and the operator from damage or injury If the Bovie Smoke Shark Surgical Smoke Evacuator is overheated or if there is an electrical surge fuses will break and the Bovie Smoke Shark Surgical Smoke Evacuator will not operate No parts are user serviceable When the Service LED Light Indicator illuminates please contact Customer Technical Services ...

Страница 17: ...rned for repair a Return Material Authorization RMA number must be obtained from Customer Service before returning the Bovie Smoke Shark Surgical Smoke Evacuator Step 4 Use the original packaging material to return your Bovie Smoke Shark Surgical Smoke Evacuator whenever possible If you do not have the original packaging material ask Customer Service for advice on how to pack for return shipment S...

Страница 18: ...he return Exchange will be of an equivalent dollar value of returned merchandise less a restocking and handling fee of new unused unopened equipment or disposables EXCEPTIONS Defective merchandise may be returned for replacement only Please contact Symmetry Surgical Customer Service before shipping back merchandise Incorrectly shipped merchandise is exempt from restocking fees Please contact Symme...

Страница 19: ...ween the parties concerning the subject matter of this Agreement The terms of any purchase order invoice or similar document used to implement this Agreement shall be subject to and shall not modify this Agreement This Agreement may only be amended by a written agreement signed by both parties Warranties All warranties provided by Symmetry Surgical Inc with respect to any Product sold are as descr...

Страница 20: ...e product is used in a manner or for purposes other than intended Bovie products are exclusively distributed by Symmetry Surgical Inc 3034 Owens Drive Antioch TN 37013 USA USA Canada 800 251 3000 Fax 800 342 3272 The revision level of this manual is specified by the highest revision letter found on either the inside front cover or enclosed errata pages if any Manual Number 905214 Rev A 07 2020 Uni...