Instructions for Use

 

 

 

Boule Cleaning Kit Medonic

 

 

 

 

 

Boule Medical AB 
Domnarvsgatan 4 
SE-163 53 Spånga, Sweden 
E-mail[email protected] 
www.boule.com 

 

 

11452-4 

2019-06-18 

 

1504093 

 

Medonic M-Series 

Intended Use 

The  Boule  Cleaning  Kit  is  intended  for  use  with  the  Medonic  M-
Series  Hematology  Systems.  The  kit  contains  3  different  cleaning 
solutions  as  indicated  on  their  labels.  Using  this  cleaning  kit  at  the 
stated regular intervals will ensure: 
 

-

 

Longer lifetime of instrument tubing  

-

 

Lower risk of bacterial grow 

-

 

Lower background counts 

-

 

Increased mean time between failure (higher MTBF) 

Contents 

1.

 

Enzymatic Cleaner, 450 mL, blue-colored solution 

2.

 

Hypochlorite Cleaner; 450 mL, yellow-colored solution 

3.

 

Detergent Cleaner, 450 mL, red-colored solution 
 

The contents can be used for three full and complete cleaning cycles. 
Remaining cleaning solution should be disposed of. 

Storage and stability 

The  Boule  Cleaning  kit  can  be  transported  at  ambient  temperature 
(avoid  freezing,  allow  adaption  to  room  temperature  before  use). 
Always store the bottles in an upright position to avoid any leakage 
from the cap. 

Cleaning Interval 

It  is  recommended  to  use  the  cleaning  kit  at  the  following  time 
intervals: 
 
Less than 50 samples/day 

 

 

every 6 months 

More than 50 samples/day 

 

every 3 months 

100-200 samples/day 

 

every month 

Procedure 

It is very important that the cleaning procedure is performed in the 
correct order

 (or it may result in high PLT background counts)

The following order must be followed: 
 

1)

 

Enzymatic Cleaner (enzyme solution - blue cap) 

2)

 

Hypochlorite (2% solution - white cap) 

3)

 

Detergent Cleaner (detergent solution - red cap) 

 

Proceed as follows:

  

 

1.

 

Firstly, remove the diluent and lyse level sensors from their 
containers. 

2.

 

Press  [MENU]  on  the  touch  screen,  then  [ADVANCED], 
then  [MAINTENANCE],  and  then  [CLEANING  MENU] 
to enter the Cleaning Menu.  

3.

 

Press [CLEAN  CYCLE  EMPTY]  and  wait  for  the  system 
to empty. 

4.

 

Wet a soft tissue with the proper cleaning solution 

(e.g. the 

first  time  use  Enzymatic  Cleaner)

  and  wipe  both  level 

sensors clean on the outside. Then insert both level sensors 
directly into the cleaning solution bottle. 

5.

 

From  Cleaning  Menu  press  [CLEAN  CYCLE  FILL]. 

Allow  the  system  to  fill  up  with  the  cleaning  solution  and 
wait 

15-30 

minutes 

for 

the 

cleaner 

to 

work.

 

6.

 

Remove  both  level  sensors  and  press  [CLEAN  CYCLE 
EMPTY].  Allow the system to empty completely. 

7.

 

Repeat step 4 to 6 using the Hypochlorite (white) followed 
by the Detergent Cleaner (red). 

8.

 

Finally take a soft tissue and wet it with Diluent (

not 

Lyse) 

and wipe off any remaining cleaner from the level sensors. 
Make sure that a soft and dry tissue is used to dry the level 
sensors  from  all  remaining  liquid.  When  level  sensors  are 
clean insert both back into the original reagent containers. 
(Be  sure  not  to  interchange  the  Diluent  and  Lyse  level 
sensors.) 

9.

 

From Cleaning Menu press [CLEAN CYCLE FILL], wait 
until this cycle is completed. 

10.

 

The background count should not exceed the recommended 
values,  which  is  stated  in  the  User  manual.  Perform  a 
minimum  of  3  background  counts  before  analyzing  the 
next sample. 

Notes 

As the volume in the cleaning bottles will decrease during 
the  cleaning  cycles,  be  sure  to  secure  the  bottles  to  avoid 
tilting. 

During  the  described  cleaning  process  above,  some 
foaming  might  occur  on  the  aspiration  needle.    Remove 
foam with a soft tissue. 

In case of heavy contaminated systems, it is recommended 
to  warm  the  detergent  to  45-55  °C  prior  to  cleaning  the 
system. 

Only  the  Hypochlorite  Cleaner  has  the  ability  to  act  as  a 
disinfectant,  significantly  reducing  the  number  of  viable 
microorganisms. 

It  is  recommended  to  check  calibration  by  running  a 
control after cleaning. 

Precautions 

-

 

Boule  Hypochlorite  Cleaner  (2%)  contains  sodium 
hypochlorite  (2%).  The  solution  is  skin  irritating  and 
causes  serious  eye  irritation.  It  is  marked  with  a  hazard 
pictogram.  

The  other  two  solutions  do  not  contain 

chemicals  at  a  level  considered  to  be  hazardous  to  health. 
However, all solutions should be handled with proper care, 
and skin and eyes should be protected. Please read the SDS 
before use. Since hypochlorite solutions are oxidizing with 
a  basic  pH,  avoid  contact  with  materials  other  than  those 
indicated  in  the  instructions  (spills  give  a  remaining  smell 
on e.g. skin and stains (bleaching) on clothes etc).  

-

 

The  concentration  of  hypochlorite  decreases  with  time 
without  invalidation  of  the  functionality  within  the  given 
expiry. Close the container as soon as possible after use. 

-

 

Cleaners should only be used before stated expiration date. 
The  cleaners  should  be  disposed  of  according  to  local 
regulations.  When  draining  through  the  sewage  system, 
always use plenty of water. 

ORDERING INFORMATION AND SERVICE 

Contact your local distributor for orders and support.  Please have the 
product number ready for orders.  For other assistance contact Boule 
Medical  AB  at  phone  +46  8  7447700,  fax  +46  8  7447720  or 

[email protected]

 

For  translation  of  this  instruction  and  explanation  to  symbols  see 

www.medonic.se

/support/. 

 

Ordering no: 

Description 

 

Packaging 

1504111 

Boule Cleaning Kit 

 

3 x 450 ml

 