Boston Scientific LithoVue Empower Скачать руководство пользователя страница 1

LithoVue Empower

 TM

  

Retrieval Deployment Device

Assembly Instructions

Please refer to directions for use for intended use statement, relevant warnings, precautions, side effects and contraindications of the device.

A)  Connect to a Nitinol Basket

STEP 1

To secure the retrieval device onto 
LithoVue Empower Device, the basket 
must be in the full open position by 
sliding the handle back completely.

STEP 2

Pick up the LithoVue Empower 
Device and hold down the trigger. 

STEP 3

Snap your Boston Scientific retrieval 
device into LithoVue Empower 
Device by matching the blue 
ridge on the handle to the blue 
indentation on LithoVue Empower 
Device. The handle will snap into 
place when secured correctly.  

STEP 4

To connect LithoVue Empower Device 
to your LithoVue Ureteroscope, 
align the pin with the D/U 
markers on the ureteroscope.  

This side must be attached first for 
the device to function properly. When 
aligned correctly, pressing down 
will snap the device into place.

STEP 5

Verify that your final assembly 
matches the above image. 

STEP 6

Feed the basket through your scope 
adaptor. Ensure the adaptor is 
completely open before feeding your 
device through or use an introducer.

Please note that if you are using an 
adaptor other than a tuohy borst 
style adaptor, you MUST use an 
introducer to insert the basket into 
the working channel of the scope.

B) Connect to the LithoVue

 Ureteroscope

Отзывы: