69
LATITUDE
®
Communicator Patient Manual
English
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment—
guidance
Radiated
RF (IEC
61000-4-3)
3 V/m
80 MHz to
2.5 GHz
10 V/m
80 MHz to
1000 MHz
h
3 V/m
1 GHz to
2.7 GHz
i
8 V/m
5.1 GHz to
5.8 GHz
d = 0.5
√
P (80 MHz to 1000 MHz)
d = 0.29
√
P (380 MHz to 390 MHz)
d = 0.36
√
P (430 MHz to 470 MHz)
d = 0.21
√
P (800 MHz to 960 MHz)
d = 1.7
√
P (1 GHz to 2.7 GHz)
d = 0.21
√
P (1.7 GHz to 1.99 GHz)
d = 0.21
√
P (2.4 GHz to 2.57 GHz)
d = 0.63
√
P (5.1 GHz to 5.8 GHz)
where
P
is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation
distance in meters (m).
Field strengths from
fi
xed RF
transmitters, as determined
by an electromagnetic site
survey,
e
should be less than
the compliance level in each
frequency range.
f
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Non-iodizing
electromagnetic
radiation
a. The LATITUDE Communicator is intended for use in the electromagnetic
environment speci
fi
ed above. The customer or the user of the LATITUDE
Communicator should assure that it is used in such an environment.
b. At 80 MHz and 1 GHz, the higher frequency range applies.
c. These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and re
fl
ection from structures, objects, and people.
d. The separation distances are calculated using the following equation
adopted from the 4th edition draft of IEC 60601-1-2:
d
= (5/
E
)
√
P
where
P
is
the maximum output power of the transmitter in watts (W) according to the
transmitter manufacturer,
d
is the separation distance in meters (m), and
E
is
the compliance level.
e. The ISM (industrial, scienti
fi
c and medical) bands between 150 kHz and 80
MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to
27.283 MHz; and 40.66 MHz to 40.70 MHz. The amateur radio bands between
150 kHz and 80 MHz are 1.8 MHz to 2.0 MHz, 3.5 MHz to 4.0 MHz, 5.3 MHz
to 5.4 MHz, 7 MHz to 7.3 MHz, 10.1 MHz to 10.15 MHz, 14 MHz to 14.2 MHz,
Содержание LATITUDE
Страница 1: ...LATITUDE Communicator Comunicador LATITUDE PATIENT MANUAL MANUAL DEL PACIENTE ...
Страница 2: ......
Страница 3: ...English 1 Español 75 ...
Страница 4: ......
Страница 5: ...English LATITUDE Communicator ENGLISH ...
Страница 72: ...64 LATITUDE Communicator Patient Manual English ...
Страница 83: ......