68
LATITUDE
®
Communicator Patient Manual
English
Table 3. Guidance and manufacturer’s declaration—
electromagnetic immunity—for equipment and systems that are not
life supporting
a b c d
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment—
guidance
Portable and mobile RF
communications equipment
should be used no
closer to any part of the
LATITUDE Communicator,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distances:
Conducted
RF (IEC
61000-4-6)
3 Vrms
150 kHz
to 80 MHz
3 Vrms
10 Vrms in
ISM and
amateur
radio bands
between
150 kHz and
80 MHz
e
d
= 1.7
√
P
(150 kHz to 80 MHz)
where
P
is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation
distance in meters (m).
Field strengths from
fi
xed RF
transmitters, as determined
by an electromagnetic site
survey,
f
should be less than
the compliance level in each
frequency range.
g