LATITUDE
®
Communicator Patient Manual
English
Table 2. Guidance and manufacturer’s declaration—
electromagnetic immunity—for all equipment and systems
a b
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic
environment—guidance
Electrostatic
discharge
(ESD) (IEC
61000-4-2)
±6 kV contact
±8 kV air
±8 kV contact
±15 kV air
The relative humidity
should be at least 15
percent.
Electrical fast
transient/
burst (IEC
61000-4-4)
±2 kV for
power supply
lines
±1 kV for
input/output
lines
±2 kV for
power supply
lines
±1 kV for
input/output
lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
(IEC
61000-4-5)
±1 kV
differential
mode
±2 kV
common
mode
±1 kV
differential
mode
±2 kV
common
mode
Mains power quality
should be that of a
typical commercial or
hospital environment.