![Boston Scientific Latitude 3300 Скачать руководство пользователя страница 57](http://html.mh-extra.com/html/boston-scientific/latitude-3300/latitude-3300_operators-manual_2796085057.webp)
51
•
Do not submerge the battery in any fluids.
•
Do not connect the + and – terminals with wire or any conductive objects.
•
Do not disassemble, modify, or repair the battery.
•
Only use the Model 3300 Programmer to charge the battery. Use of any
other battery charger can permanently damage the battery or even cause
a fire or explosion.
Storing the LATITUDE Programming System
1.
Exit the current software application by pressing the End Session button.
2.
Press and release the power button
to turn off the LATITUDE
Programming System.
NOTE:
Before moving the LATITUDE Programming System, always exit
the software application and press and release the power button
to turn
off the LATITUDE Programming System, then unplug the power cord.
NOTE:
If using battery power, press and release the power button
to
turn off the device.
3.
Unplug the power cord from the wall.
4.
Unplug all equipment cables from the side panels of the LATITUDE
Programming System.
NOTE:
See each accessory’s product literature for transport and storage
conditions. Ensure each accessory is maintained within the appropriate limits.
Long Term Storage of the LATITUDE Programming System
If the Programmer is to be stored for long periods (e.g. months), remove the
battery to prevent it from discharging to a point where recharging would be
required in order to use it again. See "Battery Status, Installation,
Replacement, and Recycling" on page 45 for instructions on removing the
battery.
Maintenance Check and Safety Measures
LATITUDE Programming System Maintenance Check
Prior to each use, you should perform a visual inspection and verify the
following:
•
Mechanical and functional integrity of the LATITUDE Programming
System, cables, and accessories.
•
Legibility and adherence of the LATITUDE Programming System labels.
•
Perform "Startup" on page 24. The normal power-up process verifies that
the LATITUDE Programming System has passed its internal checks and is
ready for use.
NOTE:
The LATITUDE Programming System does not contain any user-
serviceable parts and it does not have a calibration requirement. Maintenance
does not require any additional steps.