External Display
You can use an external monitor (or equivalent) that can synchronize to any
horizontal scan frequency.
NOTE:
External monitors may require an adapter and/or cable to connect to
the DisplayPort on the Programmer.
NOTE:
Equipment connected to the external connections must comply with
applicable standards for data processing equipment and for medical
equipment.
WARNING:
The use of any cables or accessories with the LATITUDE Programming
System other than those provided by or specified by Boston Scientific could
result in increased electromagnetic emissions, decreased electromagnetic
immunity, or electrical shock of the LATITUDE Programming System. Anyone
connecting such cables or accessories to the LATITUDE Programming
System, including the use of MSOs (Multiple Socket Outlets), may be
configuring a medical system and is responsible to ensure that the system
complies with the requirements of IEC/EN 60601-1, Clause 16 for medical
electrical systems.
CONNECTIONS
Refer to Figure 2 Right Side Panel of the Programmer on page 17 and Figure 3
Left Side Panel of the Programmer on page 17 to identify the port connections
to the Programmer.