Boston Scientific FLEXTEND Скачать руководство пользователя страница 1

FLEXTEND

®

 Lead Implant Tips

Exercise the Helix Before Implant  

December 13, 2007

                        

 ©2007 Boston Scientific Corporation or its affiliates. All rights reserved.

 

 Page 1 of 3

 

 

Before implanting the lead, verify the mechanical functioning of the helix by rotating the 
terminal pin to visually observe extension and retraction of the helix mechanism. 
1.  Attach the fixation tool to the terminal pin of the lead. Squeeze the handles together and 

place the pin in the preformed groove. Release the tension on the handles to secure the 
terminal pin in the fixation tool (Figure 1). 

                     

 

  

  
 
 
 
 

Figure 1. Fixation tool attached to terminal pin 

 
2.  Rotate the terminal pin clockwise approximately 6 to 8 turns (1 second per rotation) to 

fully extend the helix.  

3.  Rotate the terminal pin counterclockwise approximately 6 to 8 turns (1 second per 

rotation) to fully retract the helix into the lead body. 

4.  Remove the fixation tool by squeezing the handles of the tool together.  
 

CAUTIONS:  

 

Do not over-extend or over-retract the helix. Exceeding the number of turns required to 
extend or retract the helix can damage the lead. 

 

Do not use the lead if the helix cannot be extended or retracted or if the helix is 
deformed. 

 

Position the Lead 

Atrial position 

Ventricular position 

                          

 

1. Advance the lead into the right atrium 

using a straight stylet. 

2. With the lead in the low right atrium, 

insert the J-shaped or a curved straight 
stylet.* 

3. Gently pull the lead/stylet combination 

at the venous entry site to ensure 
contact between the lead tip and the 
endocardium. 

 
 
 
 
 
*Two different J-shape stylets are 
provided. One has a longer reach and 
may be suitable for most patient 
anatomies.  

                                       

 

1. Advance the lead into the right atrium using a straight stylet. 
2. Advance the lead through the tricuspid valve or place the lead 

tip against the lateral atrial wall and back the curved lead 
body through the tricuspid valve. A curved stylet may 
enhance maneuverability. 

3. Use fluoroscopy (lateral position) to ensure that the lead is not 

lodged in the coronary sinus and is in the ventricle. 

4. Insert a stylet into the lead and gently push the lead/stylet 

combination at the venous entry site to ensure contact 
between the lead tip and the endocardium. 

CAUTION: 

 

If the patient has a thin apical wall, consider an alternate 
fixation site. 

 

If a conscious patient feels a sharp pain, this may be an indication 

of perforation.

 

5.

 

Minimize the application of lead tip pressure by partially 
withdrawing the stylet during lead positioning. This will 
minimize tip stiffness. 

 

 
 
 

 

BACKGROUND INFORMATION 

This article provides information to 
enhance a user’s implant 
experience with Boston Scientific’s 
extendable/retractable FLEXTEND

 

and FLEXTEND 2 leads. 
 

This article does not contain a 
complete list of implant 
instructions and is not intended 
to supersede device-specific 
labeling.

 For complete implant 

instructions and potential risks, 
please refer to the appropriate

 

Instructions For Use 

manual.  

 
Lead fixation using an extendable 
helix mechanism is one risk factor 
for cardiac perforation. Refer to the 

Product Update

 entitled 

“Implantable Pacing Leads and Risk 
of Cardiac Perforation” available on 
www.bostonscientific.com for further 
information. 

 
 
 

 
 
 

 
 
 
 

 

 

CRM PRODUCTS REFERENCED* 

FLEXTEND Models 4086 / 4087 / 4088; 

FLEXTEND 2 Models 4095 / 4096 / 4097

 

*Products referenced herein may not be approved in all 

geographies.

 

 

CRM CONTACT INFORMATION 

 

Technical Services – U.S.   

1.800.CARDIAC (227.3422) 

[email protected]

 

Technical Services – Europe 

+32 2 416 7222 

[email protected]

 

LATITUDE Clinician Support 

1.800.CARDIAC (227.3422) 

[email protected]

 

Patient Services 

1.866.484.3268 – U.S. and Canada 

001.651.582.4000 – International

 

 

 

Отзывы: