Programming the Patient
Bionic Navigator™ 3D Programming Manual
90930923-31 REV A 49 of 309
en
Increasing/Decreasing Amplitude
To increase or decrease the amplitude of the selected Area:
1.
Select and release the
button under “Amplitude” to incrementally increase the
amplitude. Select and hold the
button to increase amplitude at a faster rate.
Note:
Alternatively, the Patient Controller up button may be used.
2.
Select and release the
button under “Amplitude” to incrementally decrease the
amplitude. Select and hold the
button to decrease amplitude at a faster rate.
Note:
Alternatively, the Patient Controller down button may be used
To increase or decrease the amplitude for all areas that have stimulation turned on:
1. Select the Program Tab in the Parameters Panel.
2. Increase or decrease the amplitude as described above.
Increasing/Decreasing Pulse Width
To increase or decrease the pulse width of the selected Area:
1.
Select and release the
button under “Pulse Width” to incrementally increase the pulse
width. Select and hold the
button to increase pulse width at a faster rate.
2.
Select and release the
button under “Pulse Width” to incrementally decrease the pulse
width. Select and hold the
button to decrease pulse width at a faster rate.
Note:
The Pulse Width control is not displayed when the Program Tab in the Parameters Panel is
selected.
Increasing/Decreasing Rate
To increase or decrease the rate of the selected Area:
1.
Select and release the
button under “Rate” to incrementally increase the rate. Select
and hold the
button to increase rate at a faster pace.
2.
Select and release the
button under “Rate” to incrementally decrease the rate. Select
and hold the
button to decrease rate at a faster pace.
Note:
The Rate control is not displayed when the Program Tab in the Parameters Panel is selected.