Configuring the CP and Stimulator
Bionic Navigator™ 3D Programming Manual
90930923-31 REV A 21 of 309
en
Configuring Leads
Before programming can begin, leads must be selected, positioned, and connected to the Stimulator on
the Leads screen.
Compatible Leads:
• Infinion™ 16 xxcm 16 Contact Lead
• Infinion CX xxcm 16 Contact Lead
• Linear™ xxcm 8 Contact Lead
• Linear ST xxcm 8 Contact Lead
• Linear 3-4 xxcm 8 Contact Lead
• Linear 3-6 xxcm 8 Contact Lead
• Artisan™ 2x8 Surgical Lead
• CoverEdge™ 32 xxcm 4x8 Surgical Lead
• CoverEdge X 32 xxcm 4x8 Surgical Lead
• 8 Contact Extensions
•
2x4 Splitters
•
2x8 Splitters
• Avista™ MRI xxcm 8 Contact Lead
Note:
xx denotes length (cm)
The Precision Spectra System supports 32 contact surgical leads, and any combination of 8 contact
percutaneous, 16 contact percutaneous and 16 contact surgical leads totaling up to 32 active contacts.
The Precision Novi and Precision Montage MRI Systems support 16 contact (2x8) surgical leads,
16 contact percutaneous leads, and a combination of 8 contact percutaneous leads, totaling up to
16 active contacts.
The Avista MRI Lead is only compatible with the Precision Montage MRI System.