Bionic Navigator™ 3D Programming Manual
Bionic Navigator™ 3D Programming Manual
90930923-31 REV A 12 of 309
en
Configuring the CP and Stimulator
The Configuration screen allows you to set up the CP and Stimulator for programming. On the
Configuration screen you can:
• Establish a programming connection between the CP and the Stimulator.
• Check the status of the programming connection.
• Identify which leads are used and record where they were implanted.
• Set up the connection between the leads and the Stimulator ports.
• Run LeadSync™ to identify and apply the relative offset of leads.
• Check the impedances of each contact.
Connecting to a Stimulator
In order to program a Stimulator, you must connect the CP to that Stimulator.
1.
Select the
Configuration
Expander bar to display the Connect screen.
2.
Select the
button to create a programming connection between the desired
Stimulator and CP. The Connection Status Indicator displays green when a valid connection
is made.
Note:
After connecting the desired Stimulator and CP, a Remote Control (and in the case of an ETS,
also the ETS On/Off button) can be used to modify the Stimulator’s settings” (e.g., turn stimulation ON
or OFF). Any subsequent change to the Stimulator from the CP will transfer the settings displayed on
the CP to the Stimulator.
Caution: If the Remote Control is used to turn stimulation OFF during programming, turn the
Stimulation On/Off Button on the CP to OFF before resuming programming. This will ensure that
the CP is synchronized with the stimulation status of the Stimulator.