It is recommended to perform a system check before running measurements (typically
in the morning). Once you click on Check-Up in the system menu bar, the following
window will open.
You should:
Set-up the calibration sample (black glossy target provided) on the middle
groove of the cylinder so you can see it on the live image
Make sure that the paper strip is sitting flat on the cylinder and that there is no
finger prints or dust on it
Remove all other samples
Close door
Once completed, the following window will open: