42
Amplifier signal lies outside one of the limit values:
- With the digital multimeter (DMM), check the signal of the channel amplifier concerned at X 7 (reference point X 9) to see whether
it matches the measured value on the motherboard (at X 107, X 207 and X 307).
For performing measurements on the channel amplifiers, unscrew the analyzer part (3 screws on the back of the chopper) and
temporarily fasten it with a screw in a bore-hole in the upper edge of the device's rear wall.
!
Do not forget to replace the contact washers of the 3 screws!
- With the digital multimeter, check U
Ref
of the channel amplifier in question at X 12 to X 9. U
Ref
= 5 V + 10 mV.
- With the digital multimeter, measure the bridge diagonal voltage of the channel amplifier in question at X 8 to X 10:
U
DC
= 0 V + 100 mV.
- Measure the bridge operating point at X 8 and X 10 to X 9: U
DC
= 1.0 V bis 1.4 V.
In the event of a fault in the channel amplifier or in the stripline, replace both complete receiving chambers to Section 10.3.
- If generally low measured values are obtained at X 107, X 207 and X 307, remove the lamp and check the surface of the reflector:
if a film has formed on the lamp, clean it as described in Section 10.2.
Diagram of channel amplifiers, simplified representation
Содержание ETT 008.31
Страница 1: ...1 Repair instructions Exhaust gas analyzer ETT 008 31 ETT 008 38 ...
Страница 48: ...48 8 15 5 Sensor board component location diagram ...
Страница 51: ...51 8 17 1 Motherboard component location diagram ...
Страница 61: ...61 12 Diagrams circuit diagrams and gas route diagram 12 1 Circuit diagram 1 ...
Страница 62: ...62 12 2 Circuit diagram 2 ...
Страница 63: ...63 12 3 Gas route diagram ...
Страница 65: ...65 ...