39
8.4
RAM check error
System parameters are stored in the RAM module. During the RAM check, these system parameters are monitored to ensure that they
remain within sensible limits pre-defined by the manufacturer. If the RAM contents are outside these limits, an error message results.
A faulty RAM content may have various causes:
- A faulty powwer supply (battery)
- Replaced motherboard
- Battery replaced with the exhaust gas analyzer switched off
- Defective RAM module (IC)
- External interference signals due to inadequately shielded peripheral equipment. The connections of the temperature sensor, clamp-
on speed sensor and serial interface must be equipped with perfect shield bonding.
Check that connections comply with circuit diagram in Section 12.
The test function (see Section 4.2) allows you to determine which parameters the exhaust-gas analyzer has recognized as faulty.
Determine and correct the faulty parameters.
Press "Test
u
" button 7 x :
Code numbers which define the faulty parameter(s) appear in the HC and O2 display fields.
i
The code numbers may appear individually or as the sum total of individual code numbers. In this way, several faulty parameters
can be displayed in a single number, e.g.: 1105 = 1 + 16 + 64 + 1024.
HC display
Faulty parameters
For correction see Section:
1
HC measuring channel mean value,
small measuring range
2
HC measuring channel mean value,
large measuring range
4
HC measuring channel upper range value,
small measuring range
8
HC measuring channel upper range value,
large measuring range
16
CO measuring channel mean value,
6.3 Fine-tuning the indicating accuracy
small measuring range
32
CO measuring channel mean value,
large measuring range
64
CO measuring channel upper range value,
small measuring range
128
CO measuring channel upper range value,
large measuring range
256
CO
2
measuring channel, mean value
512
CO
2
measuring channel, upper range value
1024
HC temperature coefficient
4.1.13 Temperature compensation
2048
CO temperature coefficient
O
2
display
Faulty parameters
For correction see Section:
1
Setpoint of HC calibration
2
HC calibration factor
4
Setpoint of CO calibration
For Switzerland and GB only
8
CO calibration factor
16
Setpoint of CO
2
calibration
32
CO
2
calibration factor
64
Unoccupied
128
Unoccupied
256
CO
2
Temperature coefficient
4.1.13 Temperature compensation
512
Altitude
4.1.7 Entering the altitude
1024
Pump delivery range
4.1.9 Adjusting the pump capacity
2048
Unoccupied
Содержание ETT 008.31
Страница 1: ...1 Repair instructions Exhaust gas analyzer ETT 008 31 ETT 008 38 ...
Страница 48: ...48 8 15 5 Sensor board component location diagram ...
Страница 51: ...51 8 17 1 Motherboard component location diagram ...
Страница 61: ...61 12 Diagrams circuit diagrams and gas route diagram 12 1 Circuit diagram 1 ...
Страница 62: ...62 12 2 Circuit diagram 2 ...
Страница 63: ...63 12 3 Gas route diagram ...
Страница 65: ...65 ...