41
5.14
Display test
(Testing TFT display)
5.15
Display test geometry
image
5.16
Display test full screen
all white
This menu serves to test the TFT display.
Select the
Display test
menu and open
the
Display test
menu by double clicking
on the left-hand mouse button or using the
Enter E key on the keyboard
i
Various image tests are
performed on the TFT display of the BEA.
Select the
Geometry image
menu and
open the menu by double clicking on the
left-hand mouse button or using the Enter
E
key on the keyboard
The geometry image is generated on the
BEA’s TFT display.
If the geometry image is in order, then the
Service software image displays the
following reference:
Geometry image OK.
In the event of a fault a clear text message
is issued.
Select the
Full screen, all white
menu
and open the menu by double clicking on
the left-hand mouse button or using the
Enter
E
key on the keyboard
The geometry image is generated on
the BEA’s TFT display.
If the full screen white is in order, then the
Service software image displays the
following reference:
Full screen all white OK.
In the event of a fault a clear text message
is issued.
Содержание BEA 150
Страница 91: ...92 12 Overview 12 1 Electrical wiring Siemens analysis chamber 459716 2Ha ...
Страница 94: ...95 12 4 Hosing Siemens analysis chamber X26 X24 X25 X22 X5 LP 223 X19 459716Ht ...
Страница 96: ...97 12 6 Electrical wiring Andros analysis chamber 4597162Ht ...
Страница 99: ...100 12 9 Hosing Andros analysis chamber 459869_03Ht ...
Страница 100: ...101 12 10 Gas circuit diagram Andros analysis chamber ...