OBD error messages
Error message
Possible cause
Remedy
Timeout: The OBD module did not
B+ and B- cables not connected.
Connect B+ and B- cables.
react within the waiting time.
Interconnecting cable from control
Test interconnecting cable and
module PCB to OBD PCB not connected
ensure that it is securely
or faulty.
fastened, replace if necessary.
Interface on control module PCB faulty
Test interface according to repair
instructions, Sec. 4.4.4, replace
control module PCB.
Faulty OBD PCB
Perform self-diagnosis as per
Section 3.1.
BsTest "Loop" "TFau"
$4000
"TSchr"
B+ and B- cables not connected.
Connect B+ and B- cables.
Faulty OBD PCB.
Perform self-diagnosis as per
Section 3.1.
i
If any other error messages appear, perform trouble-shooting as described in the Repair Instructions.