34
5.4
Checking the interface for
OBD
5.5
Checking the interface for OBD
extern (without CAN-protocol)
Select the
Interfaces
menu and then
OBD
.
Remove the 10-way Micro-Match plug
from X6 of the control module PCB.
Plug the 10-way shorting plug into X6 of
the control module PCB.
Start the interface test by double-clicking
the mouse or pressing Enter
E
.
If the interface test is error-free, the
message
OK
appears in the results box.
If the interface test is unsuccessful, an
error message is displayed.
For example: B0246:
1300
.
1300
= No shorting plug inserted or pins
3 and 5 of shorting plug not
jumpered.
If any error messages other than this one
are displayed, the control module PCB
must be replaced (Sec. 10.27).
For the functional test of the OBD printed circuit board you will require
a 15-pin sliding connector (shop made, see image below).
i
This menu allows you to test the func-
tion of the OBD PCB "to the outside"
(i.e. to the vehicle). This enables you
to determine whether the fault lies in
the vehicle or in the BEA.
In the BEA Service program, open the
menu
Interface test.
In this menu, select the submenu
OBD
extern.
Start the menu
OBD external.
8
1
15
9
4
2
3
5
6
7
8
9
10
1
14
12
13
11
15
B+ rt
GND sw
K
L
BUS+
BUS-
TSTBUS-
GND
B+
sw
rt
Schleifenstecker
Lötseite
Содержание BEA 150
Страница 91: ...92 12 Overview 12 1 Electrical wiring Siemens analysis chamber 459716 2Ha ...
Страница 94: ...95 12 4 Hosing Siemens analysis chamber X26 X24 X25 X22 X5 LP 223 X19 459716Ht ...
Страница 96: ...97 12 6 Electrical wiring Andros analysis chamber 4597162Ht ...
Страница 99: ...100 12 9 Hosing Andros analysis chamber 459869_03Ht ...
Страница 100: ...101 12 10 Gas circuit diagram Andros analysis chamber ...