16
4.5.6
O
2
sensor available
4.5.7
O
2
sensor not available
4.5.8
Entering lambda coefficients
OC
V
and HC
V
Open the
O
2
sensor available
menu using
the
ou
keys and confirm with
E
or by
double-clicking the mouse.
After confirmation that the O
2
sensor is not
available, the measuring chamber's
measuring computer (SI BENCH) is re-
started. After this it jumps back to the
Set
parameters
sub menu.
i
The exhaust-gas analyzer is then once
more in the warm-up phase. Further
measurement with the Service soft-
ware is disabled for 60 seconds. A
message appears on screen.
Open the
O
2
sensor not available
menu
using the
ou
keys and confirm with
E
or
by double-clicking the mouse.
After confirmation that the O
2
sensor is not
available, the measuring chamber's
measuring computer (SI BENCH) is re-
started. After this it jumps back to the
Set
parameters
sub menu.
i
The exhaust-gas analyzer is then once
more in the warm-up phase. Further
measurement with the Service soft-
ware is disabled for 60 seconds. A
message appears on screen.
i
Lambda coefficients must also be set
in accordance with regulations speci-
fic to the country of use and must be
stipulated by the law of that country.
Lambda coefficients OC
V
and HC
V
are fuel-specific values and are re-
quired to calculate the lambda value.
Enter the values for OC
V
on the keyboard
and confirm by clicking
OK
or by press-
ing
E
.
e.g.: 0.0175.
Enter the values for HC
V
on the keyboard
and confirm by clicking OK or by press-
ing
E
.
e.g.: 1.7261.
Содержание BEA 150
Страница 91: ...92 12 Overview 12 1 Electrical wiring Siemens analysis chamber 459716 2Ha ...
Страница 94: ...95 12 4 Hosing Siemens analysis chamber X26 X24 X25 X22 X5 LP 223 X19 459716Ht ...
Страница 96: ...97 12 6 Electrical wiring Andros analysis chamber 4597162Ht ...
Страница 99: ...100 12 9 Hosing Andros analysis chamber 459869_03Ht ...
Страница 100: ...101 12 10 Gas circuit diagram Andros analysis chamber ...