2.7 Error message
When an error occurs, the error code is displayed in the display area 8 (see 1.2).
Such as error code 2 (Err2).
Err1
Measuring value (pH/mV)
exceed range
pH
:
<
-2.00 or
>
16.00
mV
:
<
-1999or
>
1999
Check that the electrode cap is
removed and that the electrode is
connected correctly and placed in the
solution to be tested. If the instrument
does not connect the electrode, insert
the shorting plug into the socket.
Err2
Measuring temperature
exceedrange
Cal
:[℃]
<
5.0 or
>
40.0
Measure
: [℃]
<
-5.0
or
>
105.0
So that the temperature of the solution
to be measured is kept within the
specified range.
Err3
Electrode zero potential is
exceed range
|Offset
| >
60mV
Please make sure the buffer solution
used is within the validity period.
Clean or replace sensor.
Err4
Electrode slope exceed range
Slope
<
70% or
>
120%
Please make sure the buffer solution
used is within the validity period.
Clean or replace sensor.
Err5
Meter can’t recognize the
buffer solution
|∆
Eref
| <
10mV
Pls make sure the buffer solution used
is within the validity period.
Check whether the same buffer is
reused during calibration.