Integrated Multi-parameter Sensor Operation manual
58
Make integrated multi-parameter analyzer immerge into Rhodamine B standard solution, and input
corresponding sample value of water at the site of “sample calibration”. Press the key of “OK”, and press
the key of “Enter”.
6.2.8 Blue-green algae calibration
Get into the interface of blue-green algae calibration, and make integrated multi-parameter analyzer
immerge into distilled water or rinsing. Press the key of “OK” at the site of “wiping before calibration”.
Press the key of “Enter” to perform cleaning procedure for one time by scratching, and eliminate the
influence of bubbles on blue-green algae calibration process. Press the key of “OK” at the site of “zero
calibration” in the interface, and press the key of “Enter”.
Make integrated multi-parameter analyzer immerge into Rhodamine B standard solution, and input
corresponding sample value of water at the site of “sample calibration”. Press the key of “OK”, and press
the key of “Enter”.
6.2.9 Fluoride ion calibration
Get into the interface fluoride ion calibration, and press the key of “calibration”. Press the key of “Enter”
to calibrate the fluoride ion. Choose first point/second point/third point/fourth point/fifth point fluoride