19
a) Steam sterilization at 121°C (250° F) for 20 minutes (placed in a
sterilization bag), high temperature sterilization should not exceed 135°C (275°
F). Steam sterilizer should follow standard EN 13060.
b) The lip hook and file clamp can be repeatedly disinfected.
6.4 Except the lip hook and file clamp mentioned above, all other parts for the
device can’t be sterilized by high temperature and pressure.
【CAUTION】
No part of the device was disinfected before leaving the factory.
7.Storage, maintenance and transportation
7.1 Storage:
a.
The product should be handled with care, away from the source, and
stored in a dry ventilated place.
b.
Not mixed with toxic, corrosive, flammable and explosive materials.
c.
The product should be stored in a space where the relative humidity is
not exceed 80%, atmospheric pressure is 70kPa, and the temperature is
-10℃~ +50℃.
7.2 Maintenance
a.
The product do not include accessories for repair usage, the repair
should be carried out by authorized person or authorized after service
center.
b.
Keep the product in a dry storage condition.
c.
Do not throw, beat or shock the product.
d.
Do not smear the product with pigments.
7.3 Transportation:
a.
Excessive impact and shake should be prevented in transportation. Lay it
carefully and lightly, don’t invert it.