Boekel
Round Tissue Bath
Models
14792 and 14792-2
Operating Instructions
N2400142 Rev. 0
Страница 1: ...Boekel Round Tissue Bath Models 14792 and 14792 2 Operating Instructions N2400142 Rev 0...
Страница 2: ...cking 3 3 2 Installation 3 4 Operation 3 4 1 Controls and Indicator Lamp 3 4 2 Setting the Temperature 3 4 2 Filling the Bath 4 5 Fault Diagnosis 4 7 Technical Specifications 4 8 Maintenance and Servi...
Страница 3: ...rinsic protection may be impaired Connect only to a power supply with a voltage corresponding to that on the serial number label Connect only to a power supply that provides a safety ground terminal B...
Страница 4: ...Tissue Bath on a flat and stable surface preferably away from drafts Plug power cord into a power supply that matches the voltage listed on the serial number label on the rear of the unit 4 Operation...
Страница 5: ...checked by competent person 3 Temperature continues to rise when not expected a Actual temperature is lower than set temperature b Temperature control circuit fault a Check set temperature b Have uni...
Страница 6: ...able from Boekel Model 14792 2AF Boekel part number 902 0068 Model 14792 2 2 1AF Boekel part number 902 0101 Carefully replace the pan into the body Make sure that no wires are pinched when inserting...